You can calibrate the probe without pressurizing the cell by adjusting the calibration gas
flow rate downward before calibration. For instance, say the process is at 3%, and the first
calibration gas is 8%. Adjust the flow of cal gas downward until the reading begins to
migrate from 8% to lower values, indicating that the process gases are now mixing with the
calibration gases.
Adjust the flow rate back up until this mixing is just eliminated. Calibrate at this flow rate.
Replace the diffuser at the first opportunity.
WARNING!
ELECTRIC SHOCK
Install all protective equipment covers and safety ground leads after troubleshooting. Failure
to install covers and ground leads could result in serious injury or death.
Troubleshooting
Reference Manual
95
Содержание Rosemount 6888A
Страница 1: ...Reference Manual 00809 0100 4890 Rev AA July 2018 Rosemount 6888A O2 Combustion Flue Gas Transmitter ...
Страница 4: ......
Страница 8: ...Contents iv Rosemount 6888A ...
Страница 61: ...Calibration Defaults Figure 3 1 Configuration startup and operation Reference Manual 53 ...
Страница 62: ...HART Menu Trees Rosemount 6888Xi Figure 3 2 Configuration startup and operation 54 Rosemount 6888A ...
Страница 63: ...Configuration startup and operation Reference Manual 55 ...
Страница 64: ...Configuration startup and operation 56 Rosemount 6888A ...
Страница 65: ...Configuration startup and operation Reference Manual 57 ...
Страница 66: ...Configuration startup and operation 58 Rosemount 6888A ...
Страница 67: ...Configuration startup and operation Reference Manual 59 ...
Страница 68: ...Configuration startup and operation 60 Rosemount 6888A ...
Страница 69: ...HART Menu Trees 375 475 Field Communicator Figure 3 3 Configuration startup and operation Reference Manual 61 ...
Страница 70: ...Configuration startup and operation 62 Rosemount 6888A ...
Страница 71: ...Configuration startup and operation Reference Manual 63 ...
Страница 72: ...Configuration startup and operation 64 Rosemount 6888A ...
Страница 74: ...Configuration startup and operation 66 Rosemount 6888A ...
Страница 75: ...Configuration startup and operation Reference Manual 67 ...
Страница 76: ...Configuration startup and operation 68 Rosemount 6888A ...
Страница 77: ...Configuration startup and operation Reference Manual 69 ...
Страница 79: ...Configuration startup and operation Reference Manual 71 ...
Страница 80: ...Configuration startup and operation 72 Rosemount 6888A ...
Страница 81: ...Configuration startup and operation Reference Manual 73 ...
Страница 82: ...Configuration startup and operation 74 Rosemount 6888A ...
Страница 93: ...D A Trim Method Simplified Figure 3 7 Configuration startup and operation Reference Manual 85 ...
Страница 94: ...Configuration startup and operation 86 Rosemount 6888A ...
Страница 104: ...Troubleshooting 96 Rosemount 6888A ...
Страница 120: ...Maintenance and service 112 Rosemount 6888A ...
Страница 128: ...Optional accessories 120 Rosemount 6888A ...
Страница 132: ...Rosemount 6888 product certifications 124 Rosemount 6888A ...
Страница 135: ...Safety data Reference Manual 127 ...