Essential instructions

Read this page before proceeding!

Emerson designs, manufactures, and tests its products to meet many national and international standards. Because these
instruments are sophisticated technical products, you must properly install, use, and maintain them to ensure they continue to
operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program
when installing, using, and maintaining Emerson products. Failure to follow the proper instructions may cause any one of the
following situations to occur: loss of life, personal injury, property damage, damage to this instrument, and warranty invalidation.

Read all instructions prior to installing, operating, and servicing the product.

If this Reference Manual is not the correct one, call 1-800-999-9307 to request the correct Reference Manual. Save this
Reference Manual for future reference.

If you do not understand any of the instructions, contact your Emerson representative for clarification.

Follow all warnings, cautions, and instructions marked on and supplied with the product.

Inform and educate your personnel in the proper installation, operation, and maintenance of the product.

Install equipment as specified in the installation instructions of the appropriate Reference Manual and per applicable local and
national codes. Connect all products to the proper electrical and pressure sources.

To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.

When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount.
Unauthorized parts and procedures can affect the product's performance, place the safe operation of your process at risk, and
may result in fire, electrical hazards, or improper operation.

Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed
by qualified people, to prevent electrical shock and personal injury.

2

Содержание Rosemount 550pH

Страница 1: ...Reference Manual 00809 0100 3550 Rev AA March 2019 Rosemount 550pH Single use pH Sensor for Bioprocessing Applications ...

Страница 2: ...rence If you do not understand any of the instructions contact your Emerson representative for clarification Follow all warnings cautions and instructions marked on and supplied with the product Inform and educate your personnel in the proper installation operation and maintenance of the product Install equipment as specified in the installation instructions of the appropriate Reference Manual and...

Страница 3: ...sor 9 Chapter 3 Wire the sensor 13 Chapter 4 Calibrate 17 4 1 One point standardization based on a grab sample 17 Chapter 5 Troubleshoot 19 5 1 Faulty reading 19 5 2 Label information too small 19 Chapter 6 China RoHS table 21 Chapter 7 EU Declaration of Conformity 23 Reference Manual Contents 00809 0100 3550 March 2019 Reference Manual 3 ...

Страница 4: ...Contents Reference Manual March 2019 00809 0100 3550 4 Rosemount 550pH ...

Страница 5: ...units Operating temperature range 39 to 104 F 4 to 40 C Pressure range 0 to 30 Psig 0 to 2 bar pH range 2 to 12 Sensor drift Less than 0 005 pH drift per day in laboratory conditions Initial one point standardization accuracy using internal sensor buffer Less than 0 1 pH at pH 7 Shelf life 2 years at 77 F 25 C expiration date on sensor label Measuring method Potentiometry pH glass electrode and do...

Страница 6: ...units Electrical connector Variopol VP 8 0 Sensitivity Greater than 57 mV pH at 77 F 25 C Electrolyte Phosphate based see SDS on product website Integral temperature sensor Pt 100 Gamma irradiation Up to 50 kGy Sensor readiness Reference Manual March 2019 00809 0100 3550 6 Rosemount 550pH ...

Страница 7: ...onal drawings See Figure 1 1 for sensor dimensions Figure 1 1 Rosemount 550pH A1M22E0VP000 Dimensional Drawing A VP8 cable connection Reference Manual Sensor readiness 00809 0100 3550 March 2019 Reference Manual 7 ...

Страница 8: ...Sensor readiness Reference Manual March 2019 00809 0100 3550 8 Rosemount 550pH ...

Страница 9: ... to Rosemount transmitters 3 Use the Calibration menu on the transmitter to go to Slope offset setting Recommendation Do not fill the bag with buffer or other solution until after 3 b to confirm sensor functionality a Input the factory calibrated slope on the sensor label into the Slope at 25 C field and press ENTER Press EXIT to go back to the Calibration menu Do not touch the value in the Offset...

Страница 10: ... to complete the initial sensor standardization Note Use pH sensor prior to expiration date See Figure 2 2 4 Clip the zip tie locking the clip in place Remove the circular warning tag if it is still present See Figure 2 3 Figure 2 3 Clipped Zip Tie 5 Press the slide lock into the unlocked position The small lip of the lock should be the side that is pressed in See Figure 2 4 Install the sensor Ref...

Страница 11: ...on 6 Use the plunger assembly to push the sensor into the activated position See Figure 2 5 Figure 2 5 Activated Sensor 7 Press the slide lock into the locked position The small tab on the slide lock should be protruding out See Figure 2 6 Reference Manual Install the sensor 00809 0100 3550 March 2019 Reference Manual 11 ...

Страница 12: ...gure 2 7 Zip Tie Postrequisites After the start up is complete there is no additional work needed on the sensor unless a one point standardization is deemed necessary See One point standardization based on a grab sample for this procedure using an off line sample Install the sensor Reference Manual March 2019 00809 0100 3550 12 Rosemount 550pH ...

Страница 13: ...See Figure 3 1 to wire the Rosemount 550pH Sensor to the Rosemount 56 1056 or 1057 Transmitter Figure 3 1 Wire the Rosemount 550pH Sensor to the Rosemount 56 1056 or 1057 Transmitter 2 See Figure 3 2 to wire the Rosemount 550pH Sensor to the Rosemount 1066 Transmitter Reference Manual Wire the sensor 00809 0100 3550 March 2019 Reference Manual 13 ...

Страница 14: ...Rosemount 550pH Sensor to the Rosemount 1066 Transmitter 3 See Figure 3 3 to wire the Rosemount 550pH Sensor to the Rosemount 5081 Transmitter Wire the sensor Reference Manual March 2019 00809 0100 3550 14 Rosemount 550pH ...

Страница 15: ...Figure 3 3 Wire the Rosemount 550pH Sensor to the Rosemount 5081 Transmitter Reference Manual Wire the sensor 00809 0100 3550 March 2019 Reference Manual 15 ...

Страница 16: ...Wire the sensor Reference Manual March 2019 00809 0100 3550 16 Rosemount 550pH ...

Страница 17: ...ndardize grab in the Calibration menu 4 Once the pH reading of the sensor is stabilized press ENTER to pass the installation prompt 5 Input the pH of the grab sample into the pH of grab sample field and press ENTER 6 Press EXIT to complete the calibration The pH reading on the screen should now equal the pH of the grab sample Note pH is a temperature dependent attribute Please ensure off line samp...

Страница 18: ...Calibrate Reference Manual March 2019 00809 0100 3550 18 Rosemount 550pH ...

Страница 19: ...ithin sensor sleeve 90 degrees 2 Repeat sensor standardization procedure 5 2 Label information too small The serial number Cal pH or slope are too small to read Recommended action Scan the two dimensional bar code on the sensor label to retrieve information Reference Manual Troubleshoot 00809 0100 3550 March 2019 Reference Manual 19 ...

Страница 20: ...Troubleshoot Reference Manual March 2019 00809 0100 3550 20 Rosemount 550pH ...

Страница 21: ...6 China RoHS table Reference Manual China RoHS table 00809 0100 3550 March 2019 Reference Manual 21 ...

Страница 22: ...China RoHS table Reference Manual March 2019 00809 0100 3550 22 Rosemount 550pH ...

Страница 23: ...7 EU Declaration of Conformity Reference Manual EU Declaration of Conformity 00809 0100 3550 March 2019 Reference Manual 23 ...

Страница 24: ...EU Declaration of Conformity Reference Manual March 2019 00809 0100 3550 24 Rosemount 550pH ...

Страница 25: ...Reference Manual 00809 0100 3550 March 2019 Reference Manual 25 ...

Страница 26: ...ST AND AFRICA Emerson Automation Solutions Emerson FZE Jebel Ali Free Zone Dubai United Arab Emirates P O Box 17033 T 971 4 811 8100 F 971 4 886 5465 liquid csc emerson com ASIA PACIFIC Emerson Automation Solutions 1 Pandan Crescent Singapore 128461 Singapore T 65 777 8211 F 65 777 0947 liquid csc emerson com Linkedin com company Emerson Automation Solutions twitter com rosemount_news Facebook com...