4
Calibrate
4.1
One-point standardization based on a grab
sample
Complete the following steps to perform a one-point calibration against an off-line
sample. This procedure is written for a Rosemount
™
56 transmitter.
Procedure
1. Press
ENTER
on the transmitter.
2. Select "S1 Measurement" or "S2 Measurement", depending on which sensor you
wish to standardize.
3. Select "Standardize (grab)" in the
Calibration
menu.
4. Once the pH reading of the sensor is stabilized, press
ENTER
to pass the installation
prompt.
5. Input the pH of the grab sample into the "pH of grab sample" field and press
ENTER
.
6. Press
EXIT
to complete the calibration.
The pH reading on the screen should now equal the pH of the grab sample.
Note
pH is a temperature dependent attribute. Please ensure off-line sample is at the same
temperature as the process, so that no errors in standardization occur.
Reference Manual
Calibrate
00809-0100-3550
March 2019
Reference Manual
17
Содержание Rosemount 550pH
Страница 4: ...Contents Reference Manual March 2019 00809 0100 3550 4 Rosemount 550pH ...
Страница 8: ...Sensor readiness Reference Manual March 2019 00809 0100 3550 8 Rosemount 550pH ...
Страница 16: ...Wire the sensor Reference Manual March 2019 00809 0100 3550 16 Rosemount 550pH ...
Страница 18: ...Calibrate Reference Manual March 2019 00809 0100 3550 18 Rosemount 550pH ...
Страница 20: ...Troubleshoot Reference Manual March 2019 00809 0100 3550 20 Rosemount 550pH ...
Страница 21: ...6 China RoHS table Reference Manual China RoHS table 00809 0100 3550 March 2019 Reference Manual 21 ...
Страница 22: ...China RoHS table Reference Manual March 2019 00809 0100 3550 22 Rosemount 550pH ...
Страница 24: ...EU Declaration of Conformity Reference Manual March 2019 00809 0100 3550 24 Rosemount 550pH ...
Страница 25: ...Reference Manual 00809 0100 3550 March 2019 Reference Manual 25 ...