20
MODEL 54eA
SECTION 5.0
SOFTWARE CONFIGURATION
TABLE 5-1. Program Settings List
(continued)
ITEM
CHOICES
FACTORY SETTINGS
CONFIGURE (continued)
D. pH measurement (Section 5.8)
1. pH sensor
Enable or disable
Enable
2. pH compensation (free chlorine only)
Auto or manual
Auto
3. pH value
0.00 - 14.00
7.00
4. pH diagnostic
On or off
Off
a. Glass impedance high
0-2000 M
Ω
(0 disables)
1000 M
Ω
b. Glass impedance low
0-900 M
Ω
(0 disables)
20 M
Ω
c. Reference impedance high
0-140 k
Ω
(0 disables)
140 k
Ω
d. Zero offset
0-999 mV
60 mV
e. Impedance temp. compensation
On or off
On
5. pH calibration
a. Auto calibration (buffer list)
Manual, Standard, Merck, Ingold, DIN 19267
Standard
b. Stabilize pH
0.01 - 0.50
0.1
c. Stabilization time
0-30 sec
10 sec
6. Solution temperature coefficient
a. Temperature coefficient
-0.044 to 0.028 pH/°C
0.000
b. Operating isopotential
-1.35 to 20.12 pH
7.00
c. Sensor isopotential
0 to 14 pH
7.00
E. Temperature compensation (Section 5.9)
1. Temperature compensation
Auto or manual
Auto
2. Manual temperature
-15 to 130°C
25°C
F. Noise Reduction (section 5.10)
Noise rejection
50 or 60 Hz
60 Hz
G. Main sensor calibration (Section 5.11)
1. Stabilize reading
a. oxygen (ppm)
0 to 20 ppm
0.05 ppm
b. oxygen (ppb)
0 to 999 ppb
50 ppb
c. oxygen (% saturation)
0 to 200%
1%
d. chlorine, monochloramine, or ozone
0 to 9999 ppm
0.05 ppm
2. Stabilize time
0 - 30 sec
10 sec
3. Sensor zero stabilization value
4. Dual range calibration
Enable or disable
disable
5. Salinity (parts per thousand)
0.0 - 99.9
o
/
oo
0.0
o
/
oo
H. Barometric Pressure (Section 5.12)
1. Barometric pressure (oxygen only)
Auto or manual
Auto
2. Barometric pressure manual
0.4 - 1.2 atm (or equivalent units)
1.00 atm
3. Barometric pressure units
mm Hg, in Hg, bar, kPa
mm Hg
I. Security (Section 5.13)
1. Lock all
000-999 (000 disables)
000
2. Lock program
000-999 (000 disables)
000
3. Lock configuration
000-999 (000 disables)
000