MODEL 54eA
SECTION 5.0
SOFTWARE CONFIGURATION
5.6 CHANGING OUTPUT PARAMETERS (continued)
e.
The display returns to the “Output: Process” screen. Press Exit
(F1). The display returns to the “Output Measurement” screen.
Move the cursor to "Control mode" and press Enter (F4).
f.
Press Edit (F4). Use the
key to toggle between "Normal" and
"PID". Press Save (F4) to store the selection.
8.
Output setup for normal outputs:
a.
Move the cursor to the desired output setup and press Enter (F4).
b.
Use the
and
arrow keys to move the cursor to the desired
parameter. Press Edit (F4). Use the arrow keys to change the set-
ting to the desired value and press Save(F4) to store the value.
Range:
Choose 4-20 mA or 0-20 mA.
Dampen:
Dampening averages the output current, thus smooth-
ing out a noisy reading. Higher values provide more smoothing
but increase the response time of the output.
Hold and Fixed Hold:
If the analyzer is placed in hold, the out-
puts will either remain at the last value or go to a fixed value
selected by the user. The fixed value must be between 0 and
22.00 mA.
Fault:
If the analyzer detects a fault, the output will signal the fault
by going to a user-selected current between 0 and 22.00 mA.
For allowed values, see Table 5-1.
9.
Output setup for PID outputs:
a.
Move the cursor to the desired output setup and press Enter (F4).
b.
Use the
and
keys to move the cursor to the desired param-
eter. Press Edit (F4). Use the arrow keys to change the setting to
the desired value and press Save (F4) to store the value.
Setpoint:
Setpoint is usually the value at which the process is
being controlled. Typically, the output will be 4 mA (or 0 mA) when
the value is near the setpoint.
Proportional:
Proportional is the same as proportional band and
is the range over which control is being used. It is the opposite of
process gain. Smaller values provide tighter control.
Ctrl mode
: Normal
Exit
Edit
Range
: 4-20 mA
Dampen: 0 sec
Hold: Last Value
Exit
Edit
Output 1 Control
Output 1 Setup
Output 2 Control
Exit
Enter
Output 1 Control
Output 1 Setup
Output 2 Control
Exit
Enter
Setpoint
: 1.000 ppm
Proportional: 100.0%
Integral: 0 sec
Exit
Edit
31
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com