17
MODEL 54eA
SECTION 5.0
SOFTWARE CONFIGURATION
SECTION 5.0
SOFTWARE CONFIGURATION
The instrument is configured at the factory to measure oxygen.
Figure 5-1 is an outline of the menu structure.
Table 5-1 lists the default settings and the range of choices available for each setting. To reduce the chance of
error when configuring the controller the first time, enter settings in the order shown in the table.
TABLE 5-1. Program Settings List
Continued on the following page
ITEM
CHOICES
FACTORY SETTINGS
SETPOINTS
A. Alarms (Section 5.1)
1. Alarm 1 (low action)
a. if oxygen (ppm)
-99 to 99 ppm
0 ppm
b. if oxygen (ppb)
-999 to 999 ppb
0 ppb
c. if oxygen (% saturation)
0 to 200%
0 %
d. if chlorine, monochloramine, or ozone
-9999 to 9999 ppm
0 ppm
e. if pH
-2.00 to 20.00
0.00
f. if temperature
-5 to 130°C
0.1°C
2. Alarm 2 (high action)
a. if oxygen (ppm)
-99 to 99 ppm
20 ppm
b. if oxygen (ppb)
-999 to 999 ppb
900 ppb
c. if oxygen (% saturation)
0 to 200%
200 %
d. if chlorine, monochloramine, or ozone
-9999 to 9999 ppm
20 ppm
e. if pH
-2.00 to 20.00
14.00
f. if temperature
-5 to 130°C
130°C
3. Alarm 3
See alarm 2
See alarm 2
B. Outputs (Section 5.2 and 5.3)
1. Output 1 or 2: 4 mA setting
a. if oxygen (ppm)
-99 to 99 ppm
0 ppm
b. if oxygen (ppb)
-999 to 999 ppb
0 ppb
c. if oxygen (% saturation)
0 to 200%
0 %
d. if chlorine, monochloramine, or ozone
-9999 to 9999 ppm
0 ppm
e. if pH
-2.00 to 20.00
0.0
f. if temperature
-5 to 130°C
0.1°C
2. Output 1 or 2: 20 mA setting
a. if oxygen (ppm)
-99 to 99 ppm
20 ppm
b. if oxygen (ppb)
-999 to 999 ppb
900 ppb
c. if oxygen (% saturation)
0 to 200%
200%
d. if chlorine, monochloramine, or ozone
-9999 to 9999 ppm
20 ppm
e. if pH
-2.00 to 20.00
14.00
f. if temperature
-5 to 130°C
130°C
3. Setpoint (PID)
a. if oxygen (ppm)
-99 to 99 ppm
1.00 ppm
b. if oxygen (ppb)
-999 to 999 ppb
100 ppb
c. if oxygen (% saturation)
0 to 200%
100%
d. if chlorine, monochloramine, or ozone
-9999 to 9999 ppm
1.00 ppm
e. if pH
-2.00 to 20.00
7.00
f. if temperature
-5 to 130°C
25°C