7.7.4 Procedure using Deltav
Access: deltaV Explorer/transducer Block/Properties, pH Compensation tab
1. pH Compensation (Auto/Manual) and Preamp Location
Parameter: pH Compensation/Preamp Location (PH_CoMPENSAtIoN_ModE)
Select the pH compensation mode and preamp location from the following table:
2. If manual pH compensation was chosen,
Parameter: Manual pH Value (MANuAL_PH_VALuE)
Enter the desired manual pH value.
3. to enable/disable impedance diagnostic:
Parameter: Impedance diagnostics (ENABLE_dIAGNoStIC_FAuLt_SEtPoINt)
Enter the desired value.
4. Enter the high glass impedance fault limit. the default is 1000 M
W
.
Parameter: Glass Fault High Setpoint (GLASS_FAuLt_HIGH_SEtPoINt)
5. Enter the low glass impedance fault limit. the default is 10 M
W
.
Parameter: Glass Fault Low Setpoint (GLASS_FAuLt_LoW_SEtPoINt)
6. Enable or disable glass impedance temperature correction. Because glass impedance is a strong function of temper-
ature, correcting glass impedance for temperature effects is strongly recommended.
Parameter: Impedance temperature Compensation (IMPEdANCE_tEMPERAtuRE_CoMPENSAtIoN_PH)
7. Enter the reference offset limit. default is 60 mV.
Parameter: Zero offset Error Limit (ZERo_oFFSEt_ERRoR_LIMIt)
8. Select manual buffer calibration or automatic buffer calibration using the buffers listed in the table below. More infor-
mation on the buffers listed can be found in Section 13.1. using automatic buffer calibration is strongly recommended.
Parameter: Buffer Calibration (BuFFER_StANdARd)
9. Set the stabilize time for automatic buffer calibration. the range is 0 to 99 sec. the default is 10 sec.
Parameter: pH Stabilize time (PH_StABILIZE_tIME)
Enter the desired value.
10. Set the stabilization value for automatic buffer calibration. the range is 0.02 to 0.50 pH. the default is 0.02 pH.
Parameter: pH Stabilize Value (PH_StABILIZE_VALuE)
Enter the desired value.
Description
value
Digital Equivalent
Automatic pH Compensation with Preamp in transmitter
Auto, Int.
0
Manual pH Compensation with Preamp in transmitter
Man, Int.
1
Automatic pH Compensation with Sensor Preamp
Auto, Sensor
2
Manual pH Compensation with Sensor Preamp
Man, Sensor
3
Description
value
Digital Equivalent
Manual buffer calibration
Manual
0
Standard buffers
Std
1
dIN 19267 buffer
dIN
2
Ingold buffers
Ingold
3
Merck buffers
Merck
4
MODEL 5081-A
SECTION 7.0
PROGRAMMING