32
MODEL 5081-A
SECTION 5.0
DISPLAy AND OPERATION WITH INFRARED REMOTE CONTROLLER
5.2 INFRARED REMOTE CONTROLLER (IRC) - KEy FUNCTIONS
the infrared remote controller is used to calibrate and program the transmitter and to display diagnostic mes-
sages. See Figure 5-3 for a description of the function of the keys.
Hold the IRC within 6 feet of the transmitter, and not more than 15 degrees from the center of the display window.
FIGURE 5-3. Infrared Remote Controller.
RESET -
Press RESEt to end the cur-
rent operation and return to the main
display. Changes will Not be saved.
RESET does NOT return the transmit-
ter to factory default settings.
CAL -
Press CAL to access the calibra-
tion menu.
PROG -
Press PRoG to access the pro-
gram menu.
DIAG -
Press dIAG to read diagnostic
messages.
HOLD -
Press HoLd to access the
prompts used for turning on or off the
hold function.
ARROW KEyS -
use
é
and
ê
keys to
increase or decrease a number or to
scroll through items in a list. use the
ç
or
è
keys to move the cursor across a
number. A flashing word or numeral
shows the position of the cursor.
ENTER -
Press ENtER to move from a
submenu to the first prompt under the
submenu. Pressing ENtER also stores
changes in memory and advances to the
next prompt.
NExT -
Press NEXt to advance to the
next submenu or to leave a message
screen.
ExIT -
Press EXIt to end the current
operation. Changes are Not saved.