MODEL 5081-A
SECTION 7.0
PROGRAMMING
7.7.3 Procedure using the infrared remote controller
1. Press PRoG on the remote controller.
2. Press NEXt until the
PH
submenu appears.
On
will be flashing, indicating that the pH
measurement and automatic pH correction of free chlorine has been enabled.
to keep automatic pH correction enabled, press ENtER. Go to step 3.
to disable automatic pH correction, use
é
or
ê
to change
On
to
OFF
and press
ENtER. the
MAn
prompt appears. use the arrow keys to enter the pH of the sam-
ple. Press ENtER to save. Press RESEt to return to the main display.
3. the screen displays the
PAMP
(preamplifier) prompt. Press
é
or
ê
to toggle between
trAnS
and
SnSr
.
Press ENtER to save.
4. the screen displays the
dIAgnOStIC
submenu header. Prompts under this header
allow the user to change the reference offset and pH sensor diagnostic limits.
the default settings are:
to keep the default settings, press NEXt. Go to step 11.
to change the reference offset or to enable or make changes to the glass diagnostic
settings, press ENtER. Go to step 5.
5. the
rOFFSEt
(reference offset) prompt appears. use the arrow keys to change the
offset to the desired value in mV. Press ENtER to save.
6. the
dIAg
(diagnostics) prompt appears. Press
é
or
ê
to toggle between
OFF
(dis-
able) or
On
(enable). Press ENtER to save.
7. the
IMPtC
(glass impedance temperature correction) prompt appears. Press
é
or
ê
to toggle between
OFF
(disable) or
On
(enable). Because glass impedance is a
strong function of temperature, correcting glass impedance for temperature effects is
strongly recommended. Press ENtER to save.
8. the
GFH
(glass fault high) prompt appears. use the arrow keys to change the set-
ting to the desired value. the default setting is 1000 M
W
. Entering 0000 disables the
feature. Press ENtER to save.
When the glass electrode impedance exceeds the limit, the transmitter displays the
GLASSFAIL
diagnostic message and sets a fault condition.
9. the
GFL
(glass fault low) prompt appears. use the arrow keys to change the setting
to the desired value. the default setting is 10 M
W
. Entering 0000 disables the feature.
Press ENtER to save.
When the glass electrode impedance falls below the limit, the transmitter displays the
GLASSFAIL
diagnostic message and sets a fault condition.
PRoGRAM
rOffset
EXIt
ENtER
0
60
PRoGRAM
pH
EXIt
NEXt ENtER
On
PRoGRAM
pamp
=
EXIt
NEXt ENtER
trans
PRoGRAM
dIagnostIC
EXIt
NEXt ENtER
PRoGRAM
GfH
EXIt
ENtER
1
000
PRoGRAM
GfL
EXIt
ENtER
0
010
trAnS
Preamplifier is in the transmitter
SnSr
Preamplifier is in the sensor or in a remote junction box
reference offset
60 mV
pH sensor diagnostics off
PRoGRAM
dIag
EXIt
ENtER
Off
PRoGRAM
ImptC
EXIt
ENtER
On
53