![Emerson Rosemount 3152 Скачать руководство пользователя страница 28](http://html.mh-extra.com/html/emerson/rosemount-3152/rosemount-3152_reference-manual_104708028.webp)
Rosemount 3150 Series
Reference Manual
00809-0100-4835 Rev BE
February 2018
(1) Terminal block label artwork was updated for continuous improvement. Both labels shown in Figures 2-7a and 2-7b are valid
but the label shown in Figure 2-7a is current.
19
Installation
5. Tighten until cover and housing are fully engaged
metal-to-metal (see
). Once metal-to-metal
contact has been made, it is not necessary to tighten
the cover any further.
6. Visually inspect both covers to ensure they are
installed metal-to-metal. Visual inspection is sufficient
to ensure metal-to-metal contact, however, a gap
gauge may be used for verification if desired. When
metal-to-metal contact has been made, the
acceptable gap between cover and housing will be
less than 0.010 inch (see
NOTE
Replace the cover o-rings per the steps outlined in
Electronics Housing Reassembly
section if either cover was
installed metal-to-metal and then removed.
Figure 2-7 – Terminal Block
Assembly
(1)
Figure 2-7a – Current Terminal Block Assembly
Figure 2-7b – Former Terminal Block Assembly