6
Figure 2-1 – Swagelok
®
Compression
Fitting Detail
Dimensions are nominal in inches (mm)
Assembly
The Swagelok
®
tube fittings come completely assembled and
are ready for immediate use. Do not disassemble them before
use because dirt or foreign materials may get into the fitting
and cause leaks. Insert the tubing into the Swagelok
®
tube
fitting, make sure the tubing rests firmly on the shoulder of the
fitting and the nut is finger tight. Tighten the nut one-and-one-
quarter turns. Do not over-tighten.
Re-assembly
To reconnect, insert the tubing with pre-swaged ferrules into
the fitting until the front ferrule sits in the fitting. Tighten the
nut by hand, then rotate one-quarter turn more or to the
originally secured position. Then tighten the nut slightly.
For more detailed information regarding the specifications and
use of Swagelok
®
tube fittings, refer to:
Fittings Catalog MS-01-140
“Gaugeable Tube Fittings and Adapter Fittings”
Drain/Vent Valves
If drain/vent valves are opened to bleed process lines, torque
drain/vent valve stems to the value in