Reference Manual
00809-0100-4835 Rev BE
February 2018
Rosemount 3150 Series
Installation
12
Figure 2-4b – KTA Qualified and Design
Regions (applicable to 3152K and 3154K
models only)
Figure 2-4c – RCC-E Qualified and Design
Regions (applicable to 3153K and 3154K
models only)
2150
POWER SUPPLY (VDC)
2500
1500
500
12
1000
20
25
30
35
40
45
50
15
55
KTA
QUALIFIED
REGION
DESIGN
REGION
L
OAD (OHMS)
53
60
43
0
1900
2000
45
43
POWER SUPPLY (VDC)
2500
2000
1500
500
12
1000
20
25
30
35
40
50
15
55
RCC
QUALIFIED
REGION
DESIGN
REGION
L
OA
D
(
OH
M
S
)
53
2150
2050
60
0
38
750