7.2.8
Simulation active – Check operational mode
Alarm classification
Device status
Functional Check
Class
Warning
Possible cause
The device is in simulation mode and is not reporting actual information.
Recommended actions
1. If this behavior is not desired, stop simulation mode.
2. If the condition persists, restart the device.
Related information
Use the simulation mode
Perform a device reset
7.2.9
Device temperature over-run – Clear source of heat
Alarm classification
Device status
Out of specification
Class
Warning
Possible cause
The electronics temperature is outside the operating range.
Recommended actions
1. Verify ambient temperature is within the operating range.
2. Clear source of heat.
Related information
7.2.10
Device temperature under-run – Insulate device
Alarm classification
Device status
Out of specification
Class
Warning
Possible cause
The electronics temperature is outside the operating range.
Reference Manual
Service and troubleshooting
00809-0100-4480
September 2020
Reference Manual
49
Содержание Rosemount 1408H
Страница 6: ...Introduction Reference Manual September 2020 00809 0100 4480 6 Rosemount 1408H Level Transmitter ...
Страница 12: ...Transmitter overview Reference Manual September 2020 00809 0100 4480 12 Rosemount 1408H Level Transmitter ...
Страница 40: ...Configuration Reference Manual September 2020 00809 0100 4480 40 Rosemount 1408H Level Transmitter ...
Страница 96: ...Configuration parameters Reference Manual September 2020 00809 0100 4480 96 Rosemount 1408H Level Transmitter ...
Страница 97: ...Reference Manual 00809 0100 4480 September 2020 Reference Manual 97 ...