Reference Manual
00809-0100-4631, Rev BA
April 2007
Rosemount 1154 Series H
3-8
11. Readjust zero and span adjustments for corrected outputs:
3.964 mA at –100 inH
2
O
20.108 mA at 300 inH
2
O
There is an uncertainty of ±0.5 percent of input reading per 1,000 psi
associated with the span correction.
Zero
Zero shift with static pressure is not systematic. However, if the calibrated
range includes zero differential pressure, the effect can be trimmed out after
installation and with the unit at operating pressure.
Equalize pressure to both process connections, and turn the zero adjustment
until the ideal output at zero differential input is observed. Do not readjust the
span potentiometer.
If the transmitter does not include zero differential pressure within its
calibrated span, the zero effect or zero correction can be determined before
the unit is suppressed or elevated to eliminate the zero effect after correcting
for the span effect.
The following procedure illustrates how to eliminate the zero effect for a
non-zero differential pressure calibration. The example uses a Range Code 5
calibrated from 100 to 500 inH
2
O with 1,200 psi static line pressure.
1.
Using standard calibration procedures, calibrate the unit to the
required span, with the 4 mA or zero point corresponding to zero
differential pressure:
4 mA at 0 inH
2
O and 20 mA at 400 inH
2
O
2.
Apply static pressure to both high and low process connections with
zero differential pressure across the transmitter, and note the zero
correction (zero shift). For example, if the output reads 4.006 mA, the
zero correction is calculated as:
4.00 mA – 4.006 mA = –0.006 mA
Note the sign associated with this correction, as this result is added
when determining the final, ideal transmitter output.
3.
Remove static pressure and correct for the span effect as outlined in
the span correction procedure. Calibrate the unit to the calculated
output values. If, for example, the span correction procedure yielded
4.029 mA and 20.144 mA, calibrate the unit for:
4.029 mA at 100 inH
2
O
20.144 mA at 500 inH
2
O
4.
Add the zero correction (–0.006 mA), found in Step 2, to the ideal
zero point value calculated in Step 3.
4.029 mA + (–0.006 mA) = 4.023 mA
5.
To eliminate the zero effect, readjust the zero potentiometer so the
output reads the ideal zero point calculated in Step 4 (do not readjust
the span potentiometer). Note that all the calibration points will shift
the same amount toward the correct reading. The example output is
now 4.023 mA at 100 inH
2
O.
The transmitter output is now 4–20 mA over its calibrated span when the unit
is operated at 1,200 psi static line pressure.
Содержание Rosemount 1154DH
Страница 6: ...Reference Manual 00809 0100 4631 Rev BA April 2007 Rosemount 1154 Series H 1 2 ...
Страница 10: ...Reference Manual 00809 0100 4631 Rev BA April 2007 Rosemount 1154 Series H TOC 4 ...
Страница 12: ...Reference Manual 00809 0100 4631 Rev BA April 2007 Rosemount 1154 Series H 1 2 ...
Страница 52: ...Reference Manual 00809 0100 4631 Rev BA April 2007 Rosemount 1154 Series H 6 10 ...
Страница 55: ......