Reference Manual
00809-0100-4302, Rev BA
January 2008
5-7
Rosemount 1153 Series B
Electrical Housing
Reassembly
1.
Replace the zero and span adjustment screw O-rings (19) whenever
you remove the zero and span adjustment screws (16). Lightly grease
the O-rings with Dow Corning 55 Silicone O-ring Grease (Rosemount
P/N 01153-0248-0001 or P/N 01153-0053-0001). Reinstall the
adjustment screws and secure with snap rings (18).
2.
Align the zero and span adjustment screws with the potentiometer
stems on the calibration board (5) and push the calibration board onto
the bayonet pins.
3.
Slide the header assembly board (4) onto the bayonet pins with the
component side toward the pins. Slide any excess wire behind the
calibration board,
taking care to avoid kinks
.
4.
Push the amplifier board (6) onto the bayonet pins and secure with
holding screws (7). Use nominal torque of 10 in-lb (1.1 N-m).
5.
Carefully replace the cover and tighten to 16.5 ft-lb (22.4 N-m)
(See “Preliminary” on page 5-6).
6.
Replace the nameplate (17), and secure with two nameplate
screws (20).
Содержание Rosemount 1153 Series B
Страница 6: ...Reference Manual 00809 0100 4302 Rev BA January 2008 Rosemount 1153 Series B iv ...
Страница 9: ...Reference Manual 00809 0100 4302 Rev BA January 2008 TOC 3 Rosemount 1153 Series B ...
Страница 10: ...Reference Manual 00809 0100 4302 Rev BA January 2008 Rosemount 1153 Series B TOC 4 ...
Страница 12: ...Reference Manual 00809 0100 4302 Rev BA January 2008 Rosemount 1153 Series B 1 2 ...
Страница 50: ...Reference Manual 00809 0100 4302 Rev BA January 2008 Rosemount 1153 Series B 5 12 ...
Страница 62: ...Reference Manual 00809 0100 4302 Rev BA January 2008 Rosemount 1153 Series B Index 2 ...
Страница 63: ......