Reference Manual
00809-0100-4302, Rev BA
January 2008
Rosemount 1153 Series B
5-2
TEST TERMINALS
The test terminals are connected across a diode through which the loop
signal current passes. The indicating meter or test equipment shunts the
diode when connected to the test terminals. As long as the voltage across the
terminals is kept below the diode threshold voltage, no current passes
through the diode. To ensure that there is no current leaking through the diode
while making a test reading or when connecting an indicating meter, the
resistance of the test connection or meter should not exceed 10
Ω
.
BOARD CHECKOUT
NOTE
Numbers in parentheses refer to item numbers in Figure 5-2 on page 5-4.
NOTE
The Rosemount 1153 Series B Pressure Transmitter contains electronic
circuit boards which may be static sensitive.
You can easily check the printed circuit boards (5 and 6) for a malfunction by
substituting spare boards into the circuit. If this procedure turns up a
malfunctioning board, return the defective board to Rosemount Nuclear
Instruments, Inc. for replacement. Because of parts traceability, qualification
becomes the responsibility of the customer in the event of unauthorized board
repairs.
The “R” and “P” electronics circuit boards are interchangeable, and either may
be used. They must, however, be used as a pair (i.e., you must use an “R”
calibration board with the “R” amplifier board and not with the “P” amplifier
board). This also applies to the “P” calibration board.
Figure 5-1. Header Board
Connections.
Use only the procedures and new parts specifically referenced in this
manual to ensure specification performance and certification compliance.
Unauthorized procedures or parts can render the instrument dangerous
to life, limb, or property.
Process O-rings may retain some process fluid after disassembly of
process flanges. If this fluid is determined to be contaminated, take
appropriate safety measures.
Component Side Up
Содержание Rosemount 1153 Series B
Страница 6: ...Reference Manual 00809 0100 4302 Rev BA January 2008 Rosemount 1153 Series B iv ...
Страница 9: ...Reference Manual 00809 0100 4302 Rev BA January 2008 TOC 3 Rosemount 1153 Series B ...
Страница 10: ...Reference Manual 00809 0100 4302 Rev BA January 2008 Rosemount 1153 Series B TOC 4 ...
Страница 12: ...Reference Manual 00809 0100 4302 Rev BA January 2008 Rosemount 1153 Series B 1 2 ...
Страница 50: ...Reference Manual 00809 0100 4302 Rev BA January 2008 Rosemount 1153 Series B 5 12 ...
Страница 62: ...Reference Manual 00809 0100 4302 Rev BA January 2008 Rosemount 1153 Series B Index 2 ...
Страница 63: ......