Adjustment Using Calibration Gas
Functional Tests and Adjustments • 23
7
If no further changes to the other
parameters are required, press and
hold the
button for 5 seconds to
exit the Parameter list. Upon exiting
the parameters list, the gas detector
will enter offline mode for a period of
10 minutes. This allows time for the
calibration gas to dissipate after
testing. During offline mode, the gas
detector suppresses all outputs. The
display reads
oFFL
(offline) during
10 minute timeout.
NOTE: All calibration gas
mixtures have a blend tolerance.
The tolerance will limit the actual
adjustment accuracy that is
achievable.
Step
Adjustment Using Calibration
Gas Cylinders
Table 6-2
- Bump Testing