4-20 mA transmitter output settings
(continued)
Table 1-4:
Setting
Description
Calibration
Factory calibrated: 0.0 to 1000.0 = 4-20 mA output
Underrange
Output underrange: 3.8 mA
Overrange
Display overrange: 20.5 mA
Output overrange: 20.5 mA
Accuracy
± 0.05% span ± 0.004 mA
Temperature drift
0.08 μ
A/°C max from -40°C to 75°C ambient
External loop power
supply
30 VDC maximum
Output loop resist-
ance
Power supply
Minimum
Maximum
24 VDC
10Ω
700Ω
30 VDC
100Ω
1200Ω
Note: loop-powered backlight subtracts 150
Ω
from maximum resistance
figures above.
1.6
Open collector output settings
The settings below are open collector output specifications for the Internal Display. Except
when noted all specifications apply to operation at +25°C (+77°F).
Open collector output settings
Table 1-5:
Settings
Description
Output assignment Two open collector pulse outputs Out 1 and Out 2. Individually programma-
ble for rate, total, or grand total alarms; rate, total, or grand total pulse out-
puts; or retransmitting of pulse inputs; constant timed pulse output; quadra-
ture outputs (required Out 1 and Out 2); or off.
Rating
Isolated open collector, off: 24 VDC max, on: <1 V @ 150 mA max
Alarm output
Assign to rate for high or low alarm trip point.
Assign to total or grand total for total or grand total alarms.
Alarm deadband
0-100% FS, user selectable
Alarm acknowledge Front panel ACK button resets output and screen indication.
Pulse output K-Fac-
tor (Count)
K-factor (COUNT) programmable from 0.000001 to 9999999. Rate pulses
are generated as a scaled output of the rate input with one output pulse per
K-factor (count) number of input pulses. Total and grand total pulses are
generated for every total or grand total increment selected. (e.g. K-factor
value of 100 generates one pulse every time the total is incremented by 100
units). Rate retransmission pulses one to one for input pulses, up to maxi-
mum output speed. K-factor is not used for retransmitting outputs.
Introduction
User manual
9
Содержание Daniel 1200 Series
Страница 8: ......
Страница 12: ...Contents iv LTM Internal Display ...
Страница 14: ...Plan 2 LTM Internal Display ...
Страница 33: ...Open collector output connections Figure 3 13 Product connections User manual 21 ...
Страница 34: ...Product connections 22 LTM Internal Display ...
Страница 35: ...Part II Install Install User manual 23 ...
Страница 36: ...Install 24 LTM Internal Display ...
Страница 38: ...Installation requirements and limitations 26 LTM Internal Display ...
Страница 40: ...Operate 28 LTM Internal Display ...
Страница 64: ...Setting the toggle time TIME Figure 5 20 Setup and programming 52 LTM Internal Display ...
Страница 98: ...Product advanced features 86 LTM Internal Display ...
Страница 106: ...Startup procedure 94 LTM Internal Display ...
Страница 107: ...Part IV Maintain Maintain User manual 95 ...
Страница 108: ...Maintain 96 LTM Internal Display ...
Страница 113: ...Maintenance User manual 101 ...