- 6 -
PCT-193 L3-0308
cAUTioN
• Once the setting of the dose range is correctly done please keep the dose setting key in a
safe place. You will not need it anymore, unless your doctor prescribes for you a different
EMD Serono growth hormone medication. Key will be needed for changing the battery.
• Pay careful attention to when nozzle and piston must be replaced and discarded. See section
5, “Maintenance and Care”, for information on how to change the nozzle and piston.
Attaching the nozzle and piston
F
Please refer to the Instruction for Use accompanying your nozzle and piston pack before
starting using nozzle and piston.
1. First peel open the blister pack containing the nozzle and piston assembly. Remove the
assembly.
2. With the cap in place, hold the cap and front housing in one hand and the rear housing in
the other hand. Rotate the rear housing so that the rear housing moves toward the front
housing as far as it will go. You may notice a slight click as the device engages (see figure in
the section
“Setting the Dose Range”).
3. Remove the cap, and pull back the nozzle locks against the spring as shown.
4. Holding the nozzle locks back, introduce the nozzle and piston assembly to the opening in
the cool.click™ 2 as shown.
5. Push the nozzle firmly so that the piston clicks into place in the cool.click™ 2.
6. Release the nozzle locks, and ensure that they move forward fully. You will notice a “click”.
IMPORTANT
If it doesn’t click, gently push the nozzle into the device until the nozzle locks spring forwards.
7. Check that the nozzle is correctly fitted by pulling it and making sure that it stays in the
device. Once locked into position, it is important that a nozzle is installed only once, to avoid
breakage.
cAUTioN
• Thoroughly read the Instruction for Use accompanying your nozzle and piston pack for a
proper use of nozzle and piston with your cool.click™ 2 device.
• The nozzle and piston are supplied as a pre-assembled accessory in a sterile blister pack.
Check that the expiry date (indicated by the sandglass symbol ) has not expired. Do not use
the nozzle and piston assembly if the blister pack is open or appears damaged.
• The nozzle and piston are fitted to the cool.click™ 2 together.