background image

EM TEST 

 

AutoWave 

Manual for Operation 

V 5.9.1 

8 / 45 

2.2. 

Rear view  

 

 

 

 

 

1

  Output channel 1...4 

2

  Framebus IN / OUT 

CAN port 

Ethernet port 

USB port 

GPIB / IEEE 488 port 

Fuse F1 DC 3.15A 

Fuse F2 AC 1A 

Mains 90V 

– 250V 

10 

Trigger IN1 / IN2 

11 

Trigger OUT1 / OUT2 

12 

DUT monitor 

13 

DC supply 12 

– 32V 

14 

Power on switch 

 

 

1.  Output Channel 1...4 

 

 

BNC output plug to controll external DC sources. 
 
Example: 
CH 1 : Default channel for the battery power supply. 
CH 2 : Auxiliary DC power for dips 
CH 3 : Auxiliary channel for Ford specs. 
CH 4 : Auxiliary channel for Ford specs. 

Output range:

10V 

Impedance 50

Ω 

 
 

2.  Framebus IN / OUT 

 

 

Daisy Chain bus with Sub D 15 poles male and 
female connectors.  
 
This port is used as communication and control bus 
between EM Test devices. 

 

 

Содержание AutoWave

Страница 1: ...rk of the cars testing emission and immunity The two major international standards describing test procedures to simulate different phenomena s related to the battery supply lines are ISO 16750 2 for...

Страница 2: ...9 1 2 45 EM TEST Switzerland GmbH Sternenhofstrasse 15 4153 Reinach BL1 Switzerland Phone 41 61 717 91 91 Fax 41 61 717 91 99 URL http www emtest com Copyright 2016 EM TEST Switzerland GmbH All right...

Страница 3: ...to the instrument s performance and functions The figures given in this manual may differ from those that actually appear on your display and screen Every effort has been made in the preparation of th...

Страница 4: ...net Netmask 32 3 2 11 Ethernet Gateway 32 3 2 12 Date 33 3 2 13 Time 33 3 2 14 LCD Contrast 33 3 2 15 Language 33 4 Test Equipment 34 5 Technical data 35 6 Maintenance 37 6 1 General 37 6 2 Calibratio...

Страница 5: ...hapes based on segments point to point matrix Controller with internal hard disk Extensions Modules Extension Board Extension board with 2 additional output channels Play 2 measuring input channels Re...

Страница 6: ...n status 2 LED Running The LED Running indicates the running play or measuring status 3 LED Trigger The LED Trigger indicates Manual external or remote trigger event Start of a sequence 4 Mode Pressin...

Страница 7: ...annels Depends of extension 2 or 4 channels are built in Example CH 1 Default channel for the battery power supply CH 2 Auxiliary DC power for dips CH 3 Auxiliary channel for Ford specs CH 4 Auxiliary...

Страница 8: ...itor 13 DC supply 12 32V 14 Power on switch 1 Output Channel 1 4 BNC output plug to controll external DC sources Example CH 1 Default channel for the battery power supply CH 2 Auxiliary DC power for d...

Страница 9: ...negotiates the use of a 10Mbit sec or 100Mbit sec connection Pin assignment 1 TXD 2 TXD 3 RXD 4 RXD 5 USB port USB memory port for data transfer to or from a memory stick The power contacts for USB de...

Страница 10: ...Stop Trigger input for event triggering This trigger inputs are connected directly to the DSP signal processor Input Signal Negative slope Input Function Remarks Trigger IN 1 Wave start Trigger IN 2...

Страница 11: ...ger outputs are generated from the DSP signal processor Max voltage 24V pull up Current 100mA NOTE For use the trigger out the user has to connect an external DC source for pull up the trigger signal...

Страница 12: ...or is settable in the AutoWave software and has the following function Disabled no function Stop stops immediately the wave Notify send a message to DUT Log file NOTE The output voltage after a stop e...

Страница 13: ...behavior of a voltage during a sequence 2 3 2 1 Installation or mount in a 19 Rack 1 Unlock the two knobs on the front side and pull out the drawer 2 Mount the AutoWave into the four bolts This fixati...

Страница 14: ...E B Stand alone equipment Ethernet iso control software uses the IEEE interface Ethernet is not supported by iso control When setting up the test national and international regulations regarding human...

Страница 15: ...EM TEST AutoWave Manual for Operation V 5 9 1 15 45 2 3 3 1 Wiring Devices Wiring Setup example with AutoWave UCS 200 VDS 200N LD200...

Страница 16: ...00 PFS 200 RDS 200 Supply connections from VDS200 to UCS200 and PFS200 can be wired inside the rack Note Do never connect The PFS200 output 0 10V in parallel with any AutoWave output In this case the...

Страница 17: ...r General tests at UCS output Dips and Drops at PFS 200 Ford AC CI 230 at RDS and VDS outputs Note The default connection between the two racks is Test Supply Out VDS Test Supply IN PFS Dips and Drops...

Страница 18: ...EM TEST AutoWave Manual for Operation V 5 9 1 18 45 Setup 4 example with Rack MPG 200 LD 200 PFS200 VDS 200N EFT 200 CNA 200 AutoWave for replace ARB 2714...

Страница 19: ...EM TEST AutoWave Manual for Operation V 5 9 1 19 45 2 3 3 2 Wiring examples with AMP 200 Setup 5 example with Rack LD 200N AMP 200N AutoWave UCS 200N VDS 200N...

Страница 20: ...ion V 5 9 1 20 45 Setup 7 example with Rack 1 RDS 200 RDS 200 RDS 200 AMP 200N AutoWave PFS 200N VDS 200N Rack 2 LD 200N UCS 200N Setup 8 example with Rack 1 RDS 200 RDS 200 AMP 200N AutoWave PFS 200N...

Страница 21: ...EM TEST AutoWave Manual for Operation V 5 9 1 21 45 2 3 3 3 Setup with PFM 200N100 Setup 11 example with Rack AutoWave PFM 200N100 VDS 200N...

Страница 22: ...EM TEST AutoWave Manual for Operation V 5 9 1 22 45 2 3 3 4 Setup with AutoWave and VDS 200Q and PFM 200N100 Setup 12 example with Rack AutoWave PFM 200N100 VDS 200Q 25 Q 50 External AMP 200...

Страница 23: ...EM TEST AutoWave Manual for Operation V 5 9 1 23 45 2 3 3 5 Setup with AutoWave and VDS 200Q and PFM 200N100 Setup 13 example with Rack AutoWave VDS 200Q100 External PFM 200N100...

Страница 24: ...e settings STOP button Return to welcome screen Startup 3 2 1 Mode Menu Figure 4 1 shows the handling of the Mode menu which rotates cyclic by pressing the Mode button 1 st line Menu or submenu title...

Страница 25: ...cates all used output channels The software AutoWave delivers the detailed information about the wave on each channel status blinking Running off Ready to start A selected wave will be repeated accord...

Страница 26: ...EM TEST AutoWave Manual for Operation V 5 9 1 26 45 Start Break and Restart same file Start Stop and return to WaveGenerator menu...

Страница 27: ...5 9 1 27 45 3 2 1 2 Menu Wave Recorder Functions Recording the voltage variation in the lab setup Replay of the measured data via an adequate dc source or amplifier Check of the DUT under real world...

Страница 28: ...AutoWave Manual for Operation V 5 9 1 28 45 3 2 1 3 Menu Wave Manager Functions Copy file to USB Memory stick Load file from USB Memory stick Delete file Function selection Select a file and Delete Co...

Страница 29: ...g figures show the configuration of the different parameters How to navigate in the Setup menu Figure 4 2 shows the handling of the Setup menu The small buttons inside the circle shows how to step thr...

Страница 30: ...he usual one and is done in two steps first step to setup the integer part second step to setup the fractional part The all step editor mode selects each digit from left to right and the value is pars...

Страница 31: ...z 100kHz 50kHz 50kHz 25kHz 25kHz 10kHz 10kHz 5kHz 5kHz 2 5kHz 2 5kHz 1kHz 1kHz 500Hz 500Hz 250Hz 250Hz 100Hz 100Hz 50Hz 50Hz 25Hz 25Hz 10Hz 10Hz 5Hz 5Hz 3 2 5 Input Range only with the option record S...

Страница 32: ...and continue according the user decision 3 2 8 GPIB Address GPIB Address for using the AutoWave with the software iso control Standard IEEE 488 Address 1 30 Default 18 Default address for iso control...

Страница 33: ...3 2 13 Time The time is used for mark the stored files Format HH MM SS H Hour M Minute S Second Mode 24 hours day Note When pressing Setup to exit the Time setup the display returns after few seconds...

Страница 34: ...on two different power supply system AC mains power DC supply from car battery AC input The AC mains power input is a wide range AC input that allows to operate the AutoWave in every country of the w...

Страница 35: ...ith 1kHz Square wave 20Vpp without Offset Output Impedance 50 Max Output Current 10mA Output short circuit protected Wave Forms Segment types DC voltage Sine Sine sweep log linear Damp Sine Sine Ramp...

Страница 36: ...or memory stick and ext hard disc I max 500mA CAN bus for trigger Frame bus internal system bus Storage battery Lithium battery Type CR2032 3V 235mAh 20 0 x 3 2 mm Buffer battery option Rechargeable b...

Страница 37: ...performed during the life cycle of a test equipment Only the customer knows all the details and therefore the customer needs to specify the calibration interval for his test equipment In reply to all...

Страница 38: ...dure Output channel Setting a defined voltage to the output channel and verification with a DMM 5 digit Measuring 0 00V 5 00V 10 00V 7 Delivery Groups 7 1 Basic equipment Arbitrary generator type Auto...

Страница 39: ...USB port on a desktop or laptop PC under Microsoft Windows XP Me 2000 or 98SE Based on USB 2 0 the DUB E100 extends the transfer speed of earlier USB Fast Ethernet adapters to true 10 100Mbps connecti...

Страница 40: ...ware detects an older firmware on the AutoWave 1 Press the OK button to enter the Device update window B When the user has different firmware versions to operate with the AutoWave 1 Press the Update b...

Страница 41: ...download of the new firmware to AutoWave 3 Press the Reboot button for Booting the AutoWave During the booting process the AutoWave will install the new firmware version A bar graph shows the booting...

Страница 42: ...level during the selected duration t1 Ramp Voltage ramp where the time t1 goes from 0 to 100 or 10 to 90 Square Square function with defined voltage parameters V1 and V2 offset and the square duration...

Страница 43: ...EM TEST AutoWave Manual for Operation V 5 9 1 43 45 Profile Square Ramp...

Страница 44: ...r electrical equipment for measurement control and laboratory use EMC Directive 2014 30 EU Standard s to which conformity is declared EN 61326 1 2013 Electrical equipment for measurement control and l...

Страница 45: ...EM TEST AutoWave Manual for Operation V 5 9 1 45 45 8 4 AutoWave General Diagram...

Отзывы: