maximum measurement error:
±0.05 % of the measurement
(in the measuring range −40 to +120 °C and between −10 to +55 °C
ambient temperature)
•
maximum cable length:
500 m
5.4.4 Analog Measurement Transmitter (EEx i) via ExMFE5 Board
Hazardous Area
Safe Area
enCore device
ExMFE5 board
P+
P+
SH
Analog pressure
transmitter
4
...
20 mA, EEx i
Fig. 5-9: Example – analog pressure transmitter (EEx i) via ExMFE5
Characteristics:
•
intrinsically safe, galvanically isolated input safe circuits of the
category [Ex ib Gb] II C for transmitter with 4 ...20mA interface;
active signal converter integrated of the category “associated
electrical equipment”
according to EN50020 in input
•
maximum measurement error:
±0.05 % of the measurement
(in the measuring range 4 to 20 mA and between 0 to +40 °C
ambient temperature)
•
maximum cable length:
500 m