3. Configuring the datalogger

3.1 Assign network address

Digi Device Discov-
ery
Fixed IP

For datalogger identification within a network environment, a unique network address is assigned to each 
datalogger. This address is made up of 3 pieces of information: IP address, subnet mask and default gate-
way. For the elproLOG ANALYZE and elproMONITOR programs, we recommend the use of a static IP 
address.
In order to avoid network conflicts, the network administrator must assign the addresses! These 3 pieces 
of information must be entered manually into each datalogger using the software Digi Device Discovery.
For additional installation information see the 

elproLOG ANALYZE operating instructions, SE3003E, 

or  the online help feature. For additional network information 

 see ECOLOG-NET service instruction 

IT6001A.
This manual can be found on the ELPRO homepage at www.elpro.com/Download/Data Sheets/ECO-
LOG-NET networkable datalogger.

3.1.1

Desktop installation

The relevant network address is assigned to the datalogger during this workstep. This work should be per-
formed at the place of use prior to final installation, e.g. in an office.

3.1.2

Documentation

Documentation for the implemented configurations. Log the network parameters on a status printout from 
the radio datalogger.

3.2 Installing the datalogger

3.2.1

Installation

- Mount the datalogger at the place of use in accordance with the network plan.
- Connect the network, power supply, sensors, contact inputs and alarm outputs.

3.2.2

Communication test

Check communication - PING

3.2.3

elproLOG ANALYZE

elproLOG ANALYZE

Set the parameters on the datalogger,

 5. 

Settings in elproLOG ANALYZE

 and 6. 

Threshold value and 

alarm function

.

Error messages no. 5 

 4. 

Menu

3.2.4

elproMONITOR

elproMONITOR

This program is used for the on-line measured value display. For a detailed description of the functions 
and use applications

elproMONITOR operation manual SM3031E or the on-line help.

The datalogger requires an external power supply for the operation of the LAN interface. After 
connection to the power supply, it takes approximately 1 minute until the datalogger responds 
via the network.

...make a note of the IP address on the datalogger type label. This is the simplest way to identify 
the datalogger during installation!

Содержание ECOLOG-NET LP4

Страница 1: ...ECOLOG NET LP4 EL6006Ee ECOLOG NET LP4 Operation Manual...

Страница 2: ...r normal conditions All main items of the software product comply with the operating instructions and the information in the help file The guarantee does not apply to consumables disposable batteries...

Страница 3: ...ng mode 8 2 8 Overlaying graphical data 8 3 Configuring the datalogger 9 3 1 Assign network address 9 3 1 1 Desktop installation 9 3 1 2 Documentation 9 3 2 Installing the datalogger 9 3 2 1 Installat...

Страница 4: ...10 Status and error messages 24 10 1 Datalogger display and elproLOG ANALYZE 24 10 2 Datalogger status in elproLOG ANALYZE 24 11 Maintenance 25 12 Technical data 26 12 1 Measuring ranges and Accuracy...

Страница 5: ...ct to CE marking The manufacturer guarantees that this product complies with the relevant guidelines see 13 Declarations of conformity This product must be disposed of in accordance with WEEE Waste el...

Страница 6: ...ft side of the housing and on the front a display and a membrane keyboard 1 Network connection 2 Connection for 4 Pt100 sensor 200 C 600 C 3 USB connection 4 Power supply contact inputs and alarm outp...

Страница 7: ...anteed that the loggers will function properly if the datalogger is exposed to tempera tures which exceed the specified threshold range Experience has shown that the battery freezes at approx 50 C tha...

Страница 8: ...y The datenlogger ECOLOG NET has a power save mode which switches off the display As a result mea surements are only made during the defined log interval Four small circles located in the display indi...

Страница 9: ...igned to the datalogger during this workstep This work should be per formed at the place of use prior to final installation e g in an office 3 1 2 Documentation Documentation for the implemented confi...

Страница 10: ...g the datalogger 3 2 5 Verification of the installation Check the installation and ensure that sensor positions alarm parameters and network addresses are cor rect Test connectors can be used as an ai...

Страница 11: ...rm or when the self sustaining function is selected and an alarm is registered 6 2 Window Setup of Alarm parameters Jump to menu item 5 Reset alarm display and alarm contacts as well as entry of the r...

Страница 12: ...ly Definition of the alarm thresholds 6 Threshold value and alarm function Reset alarm This function is used to acknowledge an alarm mes sage Set date and time Used to adjust the integrated datalogger...

Страница 13: ...Datalogger setup 6 1 Conditions Threshold values Alarm Delay Time Alarm output The measured value must be outside the defined tolerance range i e the measured value is higher than the maximum allowab...

Страница 14: ...nction Alarm thresholds Input fields for the lower and upper threshold values Hysteresis The Hysteresis is used to avoid fluttering of the alarm contact A change of state in the contact only occurs wh...

Страница 15: ...semiconductor switch break contact to datalogger GND Semiconductor contact alarm 2 Only reacts when there is an alarm This is a semiconductor switch make contact to datalogger GND Internal Buzzer Ser...

Страница 16: ...1 cycle time 2 Alarm is logged ECOLOG NET Alarm contact responds immediately ANALYZE 2 Alarm measurement ANALYZE Alarm protocol with 2 entries alarm on off MONITOR Represents an alarm for 7 cycle time...

Страница 17: ...alarm and therefore it is not recognized as an alarm 5 Reset of delay time by short threshold value violations ECOLOG NET Alarm contact does not react Even in the case of repeated short threshold valu...

Страница 18: ...of the contacts is checked during the update of the monitor data Status changes between two updates are not detected Both inputs are busy with dual functions In each case only one function should be...

Страница 19: ...ction 8 1 Pt100 sensors 1 4 Pt100 Connector metal lized DB9 female Part no 800504 Schema Pt100 3 wires Pt100 4 wires cable length 12m Part no ECA_PT100_SILxx Pt100 4 wires cable length 12 30m Part no...

Страница 20: ...20 EN ECOLOG NET LP4 EL6006Ee Pin assignments and connections Pt100 4 wires cable length 30m Part no 800520 CABLE LENGTH 12 1 Measuring ranges and Accuracy...

Страница 21: ...logger operating volt age 6 sc Alarm contact relay contact 7 sa Alarm contact relay contact 8 Gnd Ground 9 Gnd Ground Alarm contact relay contact Relay contact represented in dead state Connection sa...

Страница 22: ...und 5 V in External power supply logger operating volt age 6 Gnd Ground 7 Gnd Ground 8 Gnd Ground 9 Gnd Ground FOLLOW THE SAFETY AND APPLICATION INSTRUCTIONS OF THE POWER SUPPLY UNIT Contact input Ala...

Страница 23: ...or VDC 500mA Alarm Threshold value violation or cable break Alarm Contact open Datalogger does not require external power supply Semi conductor contact Use only for DC voltage DCV Switch load max 24VD...

Страница 24: ...of the batteries is displayed in the elproLOG ANALYZE status information Battery replacement 11 Maintenance Firmware The version of the firmware is documented in the status Display elproLOG ANALYZE CA...

Страница 25: ...ium 3 6V 2100mAh AM3 LR6 AA Battery An energy consumption counter is used to monitor datalogger battery life For this reason only the specific manufacturer recommended battery should be used Do not re...

Страница 26: ...1 C 0 2 C 400 1 C 500 0 C 0 1 C 0 3 C 500 1 C 600 0 C 0 1 C 0 5 C This information applies to the ECOLOG NET LP4 sensors at room temperature and without Pt100 Length Cable typical errors 50m 4x0 25mm2...

Страница 27: ...EN 27 ECOLOG NET LP4 EL6006Ee Technical data IT ES EN EN FR IT ES 12 3 Dimensional view 1 Battery for a battery exchange the rear wall has to be removed 2 Screws for the attachment of the rear wall...

Страница 28: ...ension cable for Pt100 sensor 0 5m for cable guides e g refrigerators ECA_4C_xxM_M12 Connection cable Pt100 up to 30m with M12 plug ECA_LIYCY4X0 34 4 wire connection cable with M12 plug for cable leng...

Страница 29: ...EN 29 ECOLOG NET LP4 EL6006Ee Declarations of conformity IT ES EN EN FR IT ES 13 Declarations of conformity...

Страница 30: ...er 04 03 2011 First edition replaces EL6003Bk A Gubler 21 12 2011 a Minor changes A Gubler 01 10 2012 b New Graphic A Gubler 03 06 2013 c No marking of the sensor cable required A Gubler 13 01 2014 d...

Страница 31: ...ation 9 E ECOLOG NET LP4 6 elproLOG ANALYZE 9 24 elproLOG CONFIG 9 elproLOG MONITOR 9 Environmental conditions 7 Error messages 11 24 Explosion protection 6 Extended setup 12 Extended setup 12 F Firmw...

Страница 32: ...Operation Manual ECOLOG NET LP4 EL6006Ee 11 2016 Head Office ELPRO BUCHS AG Lang ulistrasse 45 9470 Buchs SWITZERLAND e mail swiss elpro com For local representatives see www elpro com...

Отзывы: