26 - EN
ECOLOG-NET LP4
EL6006Ee
Technical data
12. Technical data
12.1 Measuring ranges and Accuracy
12.1.1 Temperature measurement
12.1.2 Error caused by cable symmetries at lengths over 30m
12.1.3 Error caused by 2 wire Pt100 sensors
12.2 Operating range
A
Range
Resolution Linearity
-200.0°C ...
-100.0°C
0.1°C
± 0.3°C
-99.9°C ...
400.0°C
0.1°C
± 0.2°C
400.1°C ...
500.0°C
0.1°C
± 0.3°C
500.1°C ...
600.0°C
0.1°C
± 0.5°C
This information applies to the ECOLOG-NET LP4 sensors at room temperature and without
Pt100.
Length
Cable
typical errors
50m
4x0.25mm
2
, shielded
± 0.1°C
max. 80m
4x0.34mm
2
, shielded
± 0.2°C
Wire
Cable
typical errors
2
2x0.366mm
2
± 0.27°C/m
Ambient temperature
-35°C ... 55°C
Protective class
IP30
Power supply
Standard application
24VDC
Alarm contact
Switching load
42VAC or VDC
500mA
Alarm output 1 and 2
Switching load
Alarm cable, max. length
24VDC
200mA
15m