Elmi CM-50 Скачать руководство пользователя страница 6

Operation

 

5

 

Operation

  Settings and adjustments of the parameters of the centrifuge are made on the turned on device with 
opened or closed lid. Do not forget to use the aerodynamic, magnetic lid with the rotor.
During operation parameter adjustments are not possible.

Braking levels (15000 rpm, full rotor)

Level                                                                                                    Seconds

1....................................................................................................................28
2....................................................................................................................20
3....................................................................................................................15
4....................................................................................................................13

5....................................................................................................................11

 When all the parameters are entered to start the centrifuge close the lid and press the          button.
If the settings were made with the lid closed then press the last set parameter button
          ,           or               and press the          button.    

Thus the centrifuge will start and the timer will begin the countdown. Last minute of countdown is displayed 
in seconds. When the set time runs out centrifuge stops, alarms sounds and lid unblocks.

Centrifuge СМ-50 supports a special PUSH mode to make your work more effective 
and a sleep mode to minimize power consumption.

PUSH mode is in the speed range before 1 or after 15. Duration of centrifugation in PUSH mode is 
determined by the duration of          button retention period, as soon as the button is released, 

the centrifuge stops with the set the braking level.

SLEEP mode - device goes into sleep mode automatically after 25 minutes of standstill.
To exit sleep mode, just press any button.

ATTENTION!  Do not forget to use the magnetic aerodynamic lid with the rotor!

1. Setting speed parameter:

Press the            button and set the desired parameter value with                    buttons.

2. Setting time parameter:

Press the            button and set the desired parameter value with                    buttons.

3. Setting brake level parameter:

Press the            button and set the desired parameter value with                    buttons.

4. Modes:

Содержание CM-50

Страница 1: ...Centrifuge CM 50 User s manual ISO 9001 2008...

Страница 2: ...and usage 3 Preparation to work 4 Control description 4 Operation 5 Comparison table of RPM to RCF and vice versa 6 Troubleshooting 7 Emergency opening of centrifuge lid 7 Safety precautions 8 Disinfe...

Страница 3: ...r RCF units blocking of lid during work sound alarm and signals sleep mode and automatic lid unblocking at the end of work Built in temperature and imbalance sensors protect the unit from overheating...

Страница 4: ...nto the grooves of the rotor then tighten the rotor nut with special key Usage always load the rotor symmetrically By minimizing the difference of weight between the filled test tubes you reduce the w...

Страница 5: ...e of faults or damage the centrifuge can be considered ready to work ATTENTION In the presence of faults and damage do not use the centrifuge without consulting a specialist ATTENTION Network socket m...

Страница 6: ...trifuge stops alarms sounds and lid unblocks Centrifuge 50 supports a special PUSH mode to make your work more effective and a sleep mode to minimize power consumption PUSH mode is in the speed range...

Страница 7: ...6 Comparison table of rounds per minute RPM to relative centrifugal force RCF and vice versa Operation...

Страница 8: ...is shown on the figure Er_td Er_bd Er_db Er_OP No display Error Overheating Rotor is blocked Engine failure Total imbalance of test tubes is more than 1 gram Lid not closed No main power connection P...