Rotor assembly and usage
3
Highspeed rotor 50.01 for 12 test tubes
with aerodynamic, hermetic lid.
Maximal rotation speed 15 000 rpm.
Maximal centrifugation force 15 000 rcf.
5 braking levels.
PUSH mode is valid.
Applied rotor
Rotor assembly and usage
Assembly
: put the rotor on the axis, note that the pins of the axis must get into the grooves of
the rotor, then tighten the rotor nut with special key.
Usage
: always load the rotor symmetrically. By minimizing the difference of weight between the
filled test tubes, you reduce the wear of mechanical parts of the centrifuge. If the overall
imbalance of test tubes exceeds 1 gram, centrifuge will not start and the display will show
ER_db - imbalance error code.
Rotor 50.01
must be used with the aerodynamic, magnetic lid which seals the contents of the
rotor, reduces noise level and the heating of test tubes.