5.

 

W

hen  finished,  turn  the  pump  OFF.  Slowly  and  carefully  open  the  Vent  on  the  manifold  to 

release the residual vacuum to prevent cross-contamination. 

 

Hint:

 

Do not release the manifold vacuum pressure by lifting up the corner of 96-well extraction 

plate. This will result in deformation of the manifold gasket and will affect the hermetic seal in 

future extractions.

  

 

6.

 

Load the wash buffer on 96-well extraction plate using multichannel pipette. 

7.

 

Turn  the  pump  ON  to  apply  the  vacuum,  press  slightly  on  the  edges  of  extraction  plate  to 

engage  the  proper  vacuum  seal.  Wait  till  the  entire  volume  passes  through  the  plate  and  is 

collected in Waste Tray. 

 

8.

 

Repeat the wash steps according to recommendations of kit producers.

 

9.

 

When  finished,  turn  the  pump  OFF.  Slowly  and  carefully  open  the  Vent  on  the  manifold  to 

release the residual vacuum to prevent cross-contamination. 

10.

 

Before  the  elution  step  remove  the  96-well  extraction  plate  and  set  aside.  Disassemble  the 

manifold and remove the Waste Tray from inside of the manifold (

or throw away the pipette lid 

with collected flow through as recommended in Hint section

).   

11.

 

Place the Acrylic spacer block or Adaptor Board on the bottom of the manifold. With the block 

or  board  sitting  on  the  on  the  bottom,  place  one  96  Deep  Well  Collection  Plate  to  collect  the 

eluate (

follow recommendations of kit producers

).  

12.

 

Assemble the manifold by placing the Upper Chamber on the top of  Lower Chamber with 96 

Deep Well Collection Plate inside of manifold. 

13.

 

Put back the 96-well extraction plate firmly on the gasket of Upper Chamber so the protruding 

column tips sit few millimeters above the 96 Deep Well Collection Plate. 

14.

 

Apply  the  Elution buffer  and  elute  your  samples  into  the  96  Deep  Well Collection Plate.  Use 

your  samples  in  succeeding  PCR  analysis  OR  perform  the  elution  step  with  centrifugation  to 

ensure maximal recovery (

follow recommendations of kit producers

).  

 

Hint:

 

You  may  also  remove  residual  droplets  on  the  bottom  of  extraction  plate  by tapping  the 

top of 96-well extraction plate before releasing the vacuum by Vent.

 

 

15.

 

W

hen  finished,  turn  the  pump  OFF.  Slowly  and  carefully  open  the  Vent  on  the  manifold  to 

release the residual vacuum to prevent cross-contamination. 

 

 

 

Vytvořeno: VIJU

Instruction for use ZEPHYRUS® Vacuum Pump Manifold System

Místo uložení: https://www.elisabeth.cz/produkt-zephyrus-vacuum-pump-manifold-system-cz.html
Tento dokument je majetkem ELISABETH PHARMACON, spol. s r. o. Je určen ke zveřejnění.

Verze: 190620-01

Soubor stažen: 24.06.2020 22:38 online uživatel

Strana 9 z 17

Содержание ZEPHYRUS ZVPMS-300

Страница 1: ...ction for use ZEPHYRUS Vacuum Pump Manifold System Místo uložení https www elisabeth cz produkt zephyrus vacuum pump manifold system cz html Tento dokument je majetkem ELISABETH PHARMACON spol s r o Je určen ke zveřejnění Verze 190620 01 Soubor stažen 24 06 2020 22 38 online uživatel Strana 1 z 17 ...

Страница 2: ...ering information 15 Warranty 16 Appendix 17 Vytvořeno VIJU Instruction for use ZEPHYRUS Vacuum Pump Manifold System Místo uložení https www elisabeth cz produkt zephyrus vacuum pump manifold system cz html Tento dokument je majetkem ELISABETH PHARMACON spol s r o Je určen ke zveřejnění Verze 190620 01 Soubor stažen 24 06 2020 22 38 online uživatel Strana 2 z 17 ...

Страница 3: ...has a thermal protection feature that automatically shuts off the pump when it becomes overheated 6 The temperature of the surface of pump is very high after use or during work please do not touch it to avoid burns 7 Do not use any lubricants which may damage the pump 8 Please discard packing material according to local related regulations 9 Operating condition of vacuum pump a Ambient temperature ...

Страница 4: ...pace to work with extractions the system easily fits to your existing laminar flow box and unlike many automatic systems is takes minimum of bench space The ZEPHYRUS Vacuum Pump Manifold System can be used with both with 96 well extraction plate but also with spin columns by inserting columns to Luer connectors on Adapter Board It is suitable for those laboratories that require saving time while w...

Страница 5: ...Free Vacuum Pump x 1 Zephyrus Vacuum Manifold UNI plate Vacuum Manifold x 1 Spin Column Adaptor Board x 1 Luer Connector 25 pk x 1 Luer Stopper 25 pk x 1 Waste Tray x 1 Allen Wrench x 1 Sealing Kit O ring x 1 Gasket x 1 x 1 Silicone Tube x 1 Instruction Manual x 1 Vytvořeno VIJU Instruction for use ZEPHYRUS Vacuum Pump Manifold System Místo uložení https www elisabeth cz produkt zephyrus vacuum pu...

Страница 6: ...0 Position Designation Position Designation 1 Power Switch 4 Vacuum Regulator 2 Muffler 5 Vent 3 Air Inlet 6 Vacuum Connector Vytvořeno VIJU Instruction for use ZEPHYRUS Vacuum Pump Manifold System Místo uložení https www elisabeth cz produkt zephyrus vacuum pump manifold system cz html Tento dokument je majetkem ELISABETH PHARMACON spol s r o Je určen ke zveřejnění Verze 190620 01 Soubor stažen 2...

Страница 7: ...RUS Vacuum Manifold UNI according to Installation A 1 3 2 Turn on the pump and make a test run block the inlet of the pump and regulate to the desired vacuum level by adjusting Vacuum Regulator on the bottom of Filter Cartridge 4 then turn the pump off 3 Connect the Silicone Tube to a Vacuum Connector of installed ZEPHYRUS Vacuum Manifold UNI and inlet of ZEPHYRUS Vacuum Pump as shown in previous ...

Страница 8: ... Extraction Plates 1 Place one unused 96 well extraction plate firmly on the gasket of Upper Chamber 2 Make sure that the vacuum is switched OFF on the vacuum pump 3 Load your samples in lysis buffer on the 96 well extraction plate using multichannel pipette Hint To utilize your 96 well extraction plate effectively when having few samples you can seal off the unused part of the plate with sticky s...

Страница 9: ...he bottom of the manifold With the block or board sitting on the on the bottom place one 96 Deep Well Collection Plate to collect the eluate follow recommendations of kit producers 12 Assemble the manifold by placing the Upper Chamber on the top of Lower Chamber with 96 Deep Well Collection Plate inside of manifold 13 Put back the 96 well extraction plate firmly on the gasket of Upper Chamber so t...

Страница 10: ... the hermetic seal in future extractions 6 Load the wash buffer onto the columns turn the pump ON to apply the vacuum press slightly on the edges of Adaptor Board to engage the proper vacuum seal Wait till the entire volume passes through the plate and is collected in Waste Tray 7 Repeat the wash steps according to recommendations of kit producers 8 When finished turn the pump OFF Slowly and caref...

Страница 11: ... residual vacuum to prevent cross contamination 1200 ml PES Vacuum Bottle optional Zephyrus Vacuum Pump Manifold System assembly diagram Vytvořeno VIJU Instruction for use ZEPHYRUS Vacuum Pump Manifold System Místo uložení https www elisabeth cz produkt zephyrus vacuum pump manifold system cz html Tento dokument je majetkem ELISABETH PHARMACON spol s r o Je určen ke zveřejnění Verze 190620 01 Soub...

Страница 12: ...Separate the top ring remove the old gasket and clean thoroughly the top ring groove 3 Put the new gasket into the groove and put the top ring back Please ensure that the bottom of the top ring groove is aligned with cover gasket 4 Put back the 12 screws and tighten them at corners slightly and lock other screws back Do not lock too tight too strong otherwise distorted top ring may cause a poor va...

Страница 13: ...accessories please clean them thoroughly and autoclave 121 C 1 bar 20 min if needed 4 The Vacuum Pump is not autoclavable Please clean the surface with pure water 75 ethanol or Isopropanol IPA 5 After finishing the experiment please keep pumping the air for at least 2 minutes to withdraw the residual steam that may be trapped inside of Vacuum Pump ducts 6 When the gasket O ring are worn off or dam...

Страница 14: ... Replace with Luer stoppers Opened vent valve Close the vent valve Cracked plates or adaptor board change for a new one Loose stoppers Re plug the connectors and stoppers Dirty or damage O ring and gasket Clean or replace it Vytvořeno VIJU Instruction for use ZEPHYRUS Vacuum Pump Manifold System Místo uložení https www elisabeth cz produkt zephyrus vacuum pump manifold system cz html Tento dokumen...

Страница 15: ...O ring gasket ZVM 300 9 Allen wrench ZVM 300 10 Vacuum regulator ZVM 300 11 1200 ml PES Vacuum Bottle ZVM 300 12 Muffler for Zephyrus Vacuum Pump ZVP 300 ZVM 300 13 96 Deep Well Collection Plate 2 2 ml h44 mm ZVM 300 14 96 Deep Well Collection Plate 2 2 ml h41 mm ZVM 300 15 96 Well Collection Plate 1 2 ml h24mm Vytvořeno VIJU Instruction for use ZEPHYRUS Vacuum Pump Manifold System Místo uložení h...

Страница 16: ...ed repair use of spare parts by anyone other than Elisabeth Pharmacon Ltd used in the manner not described in this manual damage caused by accident neglect misuse improper service natural forces or other causes not arising from defects in original material or workmanship Vytvořeno VIJU Instruction for use ZEPHYRUS Vacuum Pump Manifold System Místo uložení https www elisabeth cz produkt zephyrus va...

Страница 17: ...mp Manifold System Místo uložení https www elisabeth cz produkt zephyrus vacuum pump manifold system cz html Tento dokument je majetkem ELISABETH PHARMACON spol s r o Je určen ke zveřejnění Verze 190620 01 Soubor stažen 24 06 2020 22 38 online uživatel Strana 17 z 17 ...