7
Installation and Operation
A.
Zephyrus Vacuum Manifold UNI Installation
1.
Ensure that the surface of gasket and O-ring are free from dirt or debris. Dirty gasket and O-
ring would lead to insufficient hermetic seal.
2.
Put a Waste tray in Lower Chamber; assemble the manifold by placing the Upper Chamber on
the top of it.
3.
Put 96-well extraction plate or spin column Adaptor Board firmly on the gasket of Upper
Chamber according to the type of extraction kit.
Hint:
“
When using the Waste Tray inside of the manifold to collect the flow through buffers, the
Waste Tray and the interior parts of Upper and Lower manifold Chambers have to be washed and
cleaned thoroughly after each set of extractions to prevent sample contamination
.”
B.
Zephyrus Vacuum Pump Manifold System
1.
Install the
ZEPHYRUS® Vacuum Manifold UNI
according to Installation
A. 1~3.
2.
Turn on the pump and make a test run, block the inlet of the pump and regulate to the desired
vacuum level by adjusting Vacuum Regulator on the bottom of Filter Cartridge (4) then turn
the pump off.
3.
Connect the Silicone Tube to a Vacuum Connector of installed
ZEPHYRUS® Vacuum Manifold
UNI and inlet of
ZEPHYRUS® Vacuum Pump
as shown in previous part.
Hint:
„
To set the appropriate vacuum degree is a key point to achieve efficient extraction. We
suggest setting the system vacuum level according to extraction protocol before starting
extraction.
”
96-well Extraction Plate Mini Spin Column x 24
Maxi Spin Column x 8