12
D.
Replace the O-ring and Gasket
When O-ring and gasket are worn, the proper vacuum level cannot be achieved. Please
replace them with new replacement set that is supplied free with this system.
Both O-ring and gasket should be replaced at the same time.
I.
O-ring
1.
Remove existing O-ring from the bottom of the Upper Chamber.
2.
Ensure the new O-ring is free from dirt, debris, and particulate matter.
3.
Replace with new O-ring.
II.
Gasket
1.
Remove the 12 screws located on the bottom of Upper Chamber using supplied Allen wrench.
2.
Separate the top ring, remove the old gasket, and clean thoroughly the top ring groove.
3.
Put the new gasket into the groove and put the top ring back. Please ensure that the bottom
of the top ring groove is aligned with cover gasket.
4.
Put back the 12 screws and tighten them at corners slightly and lock other screws back. Do not
lock too tight too strong otherwise distorted top ring may cause a poor vacuum.