![Elimko E-PR-110 Series Скачать руководство пользователя страница 5](http://html.mh-extra.com/html/elimko/e-pr-110-series/e-pr-110-series_user-manual_2389131005.webp)
6
7
4.2 - Technical Specification
4.2 - Technical Specification
TC
RT
(B, E, J, K, L, N, R, S, T, U)
(Pt-50, Pt-100, Pt-500, Pt-1000, Ni-100, Ni-200, Ni-500,
Ni-1000)
0 ... 20 mA, 4 ... 20 mA, -2000 mV ... 2000 mV, 0 ... 10 V
TC :
RT :
±%0.5 of the reading value or ±1°C
±%0.5 of the reading value or ±1°C
Voltage / Current : ±%0.5 FS
Input Types
Accuracy
85 - 265 V AC; 50 - 60 Hz / 85-375 V DC
Operating Voltage
Power Consumption
30 W maximum
Data Storage Memory
Communication
8 GB Micro SD Flash
10 - 100 Mbit Ethernet, RS-485, USB Host,
802.11bgn 2.4GHz Wi-Fi®
Sampling Rate
100 ms (For all channels)
Touch Panel, USB Mouse and Keyboard connections
Screen Type
5.7" TFT LCD, 640×480 resolution, 18 bit color, Touch panel
Operating Tempreature
0°C ... 50°C
-10°C ... 85°C
Storage Temperature
Analog Input
Max. 24 Analog Inputs 16 bit
(isolation of channels 1000 V AC )
(1)
Analog Output
Configuration Ports
0 – 20 mA / 0 – 10 V can be configured.
Number of analog outputs is limited to 16
(1)
Weight
Max. 64 Digital Inputs
(1)
Digital Input
Max. 64 Digital Outputs, 24 V DC, 40 mA
(1)
Digital Output
Relay Output
Language
Turkish, English
Adjustable different authority (Max. for 5 user)
Max. 12 two wire transmitter
Front Panel: IP65, Rear Panel: IP20
Password Protection
Protection Class
Max. 18 Relay Outputs, SPST-NO; 250 V AC, 5 A
Mechanical Life > 10.000.000 operation
(The relay life depends on the usage configuration)
Electrical Life > 1.000.000 operation (1/10 load)
Transmitter Supply
(2)
Approx. 1.6 kg
(1)
Configuration dependent. Number of analog outputs, relay outputs and transmitter
supply capacity are interdependently limited (
).
See Section 4.1 Type Coding
(2)
The total number of two wire transmitters to be supplied, relay outputs and analog
outputs should not exceed 18 (A+B+C<=18) due to internal 24 V DC power limitation.
This limitation is valid only with the assumption of all the output will be active at the same
time.
For certain situations such as
, different output usage
scenarios etc, the number of outputs could be increased. Please contact your sales
provider for advise!!!
A= Number of two wire transmitters to be supplied
B= Number of Relay Outputs (number of individual outputs, not
the number of cards, each relay output card has 6 relay outputs)
C= Number of Analog Outputs (number of individual outputs, not
the number of cards, each analog output card has 4 analog outputs)
existence of spare outputs