Eli Ezer ESL-7800 Скачать руководство пользователя страница 14

 

10 

3.2  Patient position and fixation target----------------------------------------- 

1)  Positioning the patient’s head   

Have  the  patient  place  his  chin  on 
the  chin-rest  (17)  and  the  forehead 
against  the  forehead-rest  belt  (24). 
Adjust 

the 

chin-rest 

elevation 

adjustment  knob  (12)  below  the 
chin-rest  until  the  patient’s  canthus 
aligns with the horizontal mark (23) 
(Fig.21). 

 

Fig.21 

2)  Use of the fixation target 

 

 

For fixing the patient’s eyesight, just 
make him look at the fixation target 
(30) with the eye not to be examined. 
To change fixing position, move the 
lamp bar, as well as move the curved 
lever around the forehead. 

 

The  fixation  target  with  diopter 
compensation  supplies  a  dot  and 

concentric  circles  target.  Slide  the 
knob 

to 

adjust 

the 

diopter 

compensation  within  the  range  from 
-15D to +10D (Fig.22). 

     

 

Fig.22 

 

The fixation target with spot light is 
especially  for  the  patient  whose 
diopter  exceeds  -15D  (Fig.23). 
When  changing,  just  loosen  the 
fixation  screw,  replace  the  fixation 
target with the spot light source and 
refasten the fixation screw. 

 

Fig.23 

3.3  Base operation ------------------------------------------------------------------ 

1) 

Horizontal rough adjustment 

Keep the joystick (1) erect and move 
the base (4) to make the microscope 
move  on  the  horizontal  surface  to 
aim at the object roughly. (Fig.24). 

2)  Vertical adjustment 

Rotate  the  joystick  to  adjust  the 

microscope’s  height  until  it  aligns 
with  the  target.  Turn  the  joystick 
clockwise  to  raise  the  microscope 
and counter clockwise to lower it. 

3) 

Horizontal Fine adjustment     
Tilt  the  joystick  to  make  the 
microscope  move  slightly  on  the 
horizontal  surface.  While  watching 

Содержание ESL-7800

Страница 1: ......

Страница 2: ...rectly and firmly connected before using Ensure that the instrument is well grounded 3 Please pay attention to all the ratings of the electrical connecting terminal 4 Only use fuse according to the sp...

Страница 3: ...Fig 1...

Страница 4: ...n with Hruby lens optional 14 4 Maintenance 15 4 1 Replacing the illumination bulb 15 4 2 Replacing the reflecting mirror 15 4 3 Replacing the fuse 16 4 4 Replacing the chin rest paper 16 4 5 Adjustin...

Страница 5: ...e arm represent the two arms angle When the 0 on the ring relates to the short mark at one side of the operator the right eyepiece may be blocked and the side of the patient the left eyepiece 12 Chin...

Страница 6: ...ble One is for diopter adjustment to help the patient to see the target clearly while the other is an illuminated fixed spot 31 Reflecting Mirror Both long and short reflecting mirrors are provided Th...

Страница 7: ...3 2 Assembly This section of the manual describes how to assemble ESL 7800 slit lamp All parts should be taken out with great care from the packing case before assembling 2 1 Components Fig2 1...

Страница 8: ...Hruby Lens 1 I Hruby Lens Guide Plate 1 J Spare Illumination Bulb 1 K Chin rest Paper 1 L Focusing Test Rod 1 M Protection Cap 2 N Dust proof Cover 1 O Fixation Target 1 P Spare Long Reflecting Mirro...

Страница 9: ...the input voltage and frequency of the instrument according to that of the mains 2 Assembling the Worktable F To attach the worktable on the motorized instrument table please screw off four M8x20mm b...

Страница 10: ...tion arm with the screwdriver V Aim the assembly hole of the illumination arm at the brass shaft sleeve with care then put down let the shaft keeping close to the bottom surface well and simultaneousl...

Страница 11: ...tput socket of the power box Insert the plug of the input power cable H into the input socket of the power box Remove the cable clips from the bottom of the work table with screw driver U and wrap the...

Страница 12: ...onpart When the main power switch 8 of the power box is placed at I it turns on and O for turn off The main power switch should be set at the O position before connecting the input cable with the powe...

Страница 13: ...perator is an ametropia he should adjust the eyepiece diopter One eyepiece with four short reticle lines witch is usually placed at the right side helps to focus accurately after accessories being att...

Страница 14: ...target Slide the knob to adjust the diopter compensation within the range from 15D to 10D Fig 22 Fig 22 The fixation target with spot light is especially for the patient whose diopter exceeds 15D Fig...

Страница 15: ...ircular beams of light are available at full aperture 9 8 5 3 1 0 2 dia respectively With a slit image the slit height can be changed continuously from 1 to 9mm which is indicated through the display...

Страница 16: ...ig 29 Fig 29 6 Reflecting mirror Both short and long reflecting mirror are available in this instrument Use the long mirror in normal examination When the angle between the illumination part and the m...

Страница 17: ...e groove on the guide plate 4 Move the focus of the illumination light and the microscope to the patient s eye 5 Move the lever to locate the Hruby lens at the center of the vision field and near the...

Страница 18: ...otherwise the illumination may be uneven Fig 34 Attention The bulb is hot Place the lamp cap in the original position and rotate it clockwise and insert the connecting plugs Turn on the main power swi...

Страница 19: ...n the paper is depleted pull upward two fixing pins of the chin rest and place a new package of paper then fix the fixing pins again Fig 38 Fig 38 4 5 Adjusting the tightness of the slit width knob If...

Страница 20: ...rts such as chin rest bracket forehead rest belt with soft cloth dipped with soluble detergent or water sterilize with medicinal alcohol Fig 41 Attention Don t wipe with any corrosive detergent lest t...

Страница 21: ...nge the fuse P16 The bulb is not assembled properly Assemble the bulb properly P15 The filter lever is in the middle position or in the position of gray filter Set the filter lever to the correct posi...

Страница 22: ...25o C to 40 o C This instrument should be stored in a well ventilated room without corrosive gas where the relative humidity should be 10 to 80 and environment temperature 10o C to 40o C If the assem...

Страница 23: ...to horizontal Slit inclination 5o 10o 15o 20o four steps Filter piece Heat absorbing gray red free blue Illumination bulb 6V20W halogen bulb Movement base Fore and back movement 90mm Left and right m...

Страница 24: ......

Отзывы: