Eli Ezer ESL-7800 Скачать руководство пользователя страница 11

 

 

Fig11 

8) 

 

Removing  the  illumination  part 

shipping pad 

 

This  pad  is  used  to  protect  the  slit 
mechanism  of  the  illumination  part 
during shipping. 

 

Remove  the  rubber  band  and  gently 
pull the pad out (Fig.12). 

 

 

Fig.12 

 

9) 

 

Connecting plug   

 

Fig.13 

 

Insert  the  plug  on  the  top  of  the 
headrest  part  (D)  into  the  socket  of 
the 

lamp 

cap 

(26) 

on 

the 

illumination part (A) (Fig.13). 

 

Connect  the  two  plugs  below  the 
headrest part with the corresponding 
output socket of the power box.   

 

Insert  the  plug  of  the  input  power 
cable (H) into the input socket of the 
power box. . 

 

Remove  the  cable  clips  from  the 
bottom of the work table with screw 
driver  (U)  and  wrap  the  output  and 
input 

cables 

respectively, 

then 

reattach  them  to  the  bottom  of  the 
work table. (Fig14). 

 

Fig.14 

10)  Assembling  the  Hruby  lens  (I)  and 

the Hruby lens guide plate (J) 

 

Insert  the  Hruby  lens  (I)  into  the 
Hruby  lens  holder  (35)  on  the 
headrest  part.  Be  careful  not  to 
touch the lens surface    (Fig15.). 

 

Fig.15 

 

Place the Hruby lens guide plate (J) 
into the main shaft hole  of the base 
part  with  the  small  end  pointing  to 
the headrest part (Fig16). 

Содержание ESL-7800

Страница 1: ......

Страница 2: ...rectly and firmly connected before using Ensure that the instrument is well grounded 3 Please pay attention to all the ratings of the electrical connecting terminal 4 Only use fuse according to the sp...

Страница 3: ...Fig 1...

Страница 4: ...n with Hruby lens optional 14 4 Maintenance 15 4 1 Replacing the illumination bulb 15 4 2 Replacing the reflecting mirror 15 4 3 Replacing the fuse 16 4 4 Replacing the chin rest paper 16 4 5 Adjustin...

Страница 5: ...e arm represent the two arms angle When the 0 on the ring relates to the short mark at one side of the operator the right eyepiece may be blocked and the side of the patient the left eyepiece 12 Chin...

Страница 6: ...ble One is for diopter adjustment to help the patient to see the target clearly while the other is an illuminated fixed spot 31 Reflecting Mirror Both long and short reflecting mirrors are provided Th...

Страница 7: ...3 2 Assembly This section of the manual describes how to assemble ESL 7800 slit lamp All parts should be taken out with great care from the packing case before assembling 2 1 Components Fig2 1...

Страница 8: ...Hruby Lens 1 I Hruby Lens Guide Plate 1 J Spare Illumination Bulb 1 K Chin rest Paper 1 L Focusing Test Rod 1 M Protection Cap 2 N Dust proof Cover 1 O Fixation Target 1 P Spare Long Reflecting Mirro...

Страница 9: ...the input voltage and frequency of the instrument according to that of the mains 2 Assembling the Worktable F To attach the worktable on the motorized instrument table please screw off four M8x20mm b...

Страница 10: ...tion arm with the screwdriver V Aim the assembly hole of the illumination arm at the brass shaft sleeve with care then put down let the shaft keeping close to the bottom surface well and simultaneousl...

Страница 11: ...tput socket of the power box Insert the plug of the input power cable H into the input socket of the power box Remove the cable clips from the bottom of the work table with screw driver U and wrap the...

Страница 12: ...onpart When the main power switch 8 of the power box is placed at I it turns on and O for turn off The main power switch should be set at the O position before connecting the input cable with the powe...

Страница 13: ...perator is an ametropia he should adjust the eyepiece diopter One eyepiece with four short reticle lines witch is usually placed at the right side helps to focus accurately after accessories being att...

Страница 14: ...target Slide the knob to adjust the diopter compensation within the range from 15D to 10D Fig 22 Fig 22 The fixation target with spot light is especially for the patient whose diopter exceeds 15D Fig...

Страница 15: ...ircular beams of light are available at full aperture 9 8 5 3 1 0 2 dia respectively With a slit image the slit height can be changed continuously from 1 to 9mm which is indicated through the display...

Страница 16: ...ig 29 Fig 29 6 Reflecting mirror Both short and long reflecting mirror are available in this instrument Use the long mirror in normal examination When the angle between the illumination part and the m...

Страница 17: ...e groove on the guide plate 4 Move the focus of the illumination light and the microscope to the patient s eye 5 Move the lever to locate the Hruby lens at the center of the vision field and near the...

Страница 18: ...otherwise the illumination may be uneven Fig 34 Attention The bulb is hot Place the lamp cap in the original position and rotate it clockwise and insert the connecting plugs Turn on the main power swi...

Страница 19: ...n the paper is depleted pull upward two fixing pins of the chin rest and place a new package of paper then fix the fixing pins again Fig 38 Fig 38 4 5 Adjusting the tightness of the slit width knob If...

Страница 20: ...rts such as chin rest bracket forehead rest belt with soft cloth dipped with soluble detergent or water sterilize with medicinal alcohol Fig 41 Attention Don t wipe with any corrosive detergent lest t...

Страница 21: ...nge the fuse P16 The bulb is not assembled properly Assemble the bulb properly P15 The filter lever is in the middle position or in the position of gray filter Set the filter lever to the correct posi...

Страница 22: ...25o C to 40 o C This instrument should be stored in a well ventilated room without corrosive gas where the relative humidity should be 10 to 80 and environment temperature 10o C to 40o C If the assem...

Страница 23: ...to horizontal Slit inclination 5o 10o 15o 20o four steps Filter piece Heat absorbing gray red free blue Illumination bulb 6V20W halogen bulb Movement base Fore and back movement 90mm Left and right m...

Страница 24: ......

Отзывы: