4. Display assignment
5. The AZ16I in operation
5.1 Initial operation
5.1.1 Detection of measuring direction
On the magnetic tape is an arrow marker, which indicates the positive direction.
According to the following image, the arrow of the tape must show to the right.
The counting direction (+/-) can be changed by the register
P01
(see chapter 5.4).
Front view
5.1.2 Referencing
Setting the zero-point:
A new AZ16I unit shows always the absolute value of the magnetic tape
and should be calibrated to the zero point at one time. The reference value default setting in reg-
ister
P09
is
0
. To assign a zero point to an arbitrary position, move to the desired
zero-point and press the buttons
F + Set
together.
Reference value:
Alternatively an arbitrary value can be entered in
P09
, to set a demanded ref-
erence measurement by pressing
F + Set
together.
5156,9
Marker
Positive direction
(when P01 = default setting)
Magnetic Tape
battery condition direction arrow tool offset 3 (1, 2 or 3 possible)
absolute mode
sign measurement unit