PURELAB 7000 - US 

Operator Manual

Содержание PURELAB 7000 - US

Страница 1: ...PURELAB 7000 US Operator Manual...

Страница 2: ......

Страница 3: ...nal day screen 030 16 4 2 Continuous Recirculation 24 7 screen 029 16 5 MAINTENANCE 17 5 1 Replacing Composite vent filter LC136 17 5 2 Replacing Ultra Microfilter LC109 17 5 3 Replacing Labpure L1 pu...

Страница 4: ...THIS APPLIANCE MUST BE EARTHED It is essential that the electrical supply to the PURELAB 7000 is isolated before any items are changed or maintenance work performed The external isolator providing po...

Страница 5: ...nt filter 6 months 2 years LC175 Protek L1 cartridge 6 months 2 years LC177 Protek L2 cartridge 12 months 2 years LC181 E cartridge 36 months 2 years LC187 Labpure L1 cartridge 3 to 12 months 2 years...

Страница 6: ...s the unit to be quickly installed and commissioned It is recommended that a trained ELGA representative carry out the installation of the product 2 2 Commissioning The unit is supplied with the softw...

Страница 7: ...ment Date RIGHT Mute Alarm Down CAUTION If the Master Key is lost a new Key can only be programmed by an ELGA Service Engineer 3 1 User Recognition Keys Your unit is supplied with the following User k...

Страница 8: ...User Key does not prevent access to the PROCESS function in case of emergency 1 SWITCH on the main electrical supply to start the controller set up sequence This takes several seconds 2 PRESENT the Ma...

Страница 9: ...S Step 4 Audible alarm enabled disabled screen 020 This screen provides the option of either turning on the audible alarm causing it to sound while the alarm icon flashes or turning off the audible al...

Страница 10: ...purposes only and is not recommended to be selected as standard Step 8 RO water quality alarm QS1 screen 025 This screen is used to set the value at which the RO water quality alarm will sound This a...

Страница 11: ...ture falls below the set point 3 PRESS to select alarm point in increments of 1 ranging from 20 C to 50 C OR PRESS to proceed to Step 12 2 PRESS Note If the water temperature rises above 55 C in the s...

Страница 12: ...nutes 5 PRESS to enter times 6 PRESS Note Night service will only be operational if Continuous Operation is not enabled Step 14 Operational days selection screen 031 Select the days that the PURELAB 7...

Страница 13: ...unit will automatically restart after a power failure If AUTO restart is not selected the unit will stay in standby after a power failure 1 PRESS to change mode ON OR PRESS to proceed to Step 17 2 PR...

Страница 14: ...during normal operation and current data will be transmitted if data output screen 035 is selected Step 20 Programming of User Keys screen 037 During the life of the PURELAB 7000 you may need to delet...

Страница 15: ...reader 2 PRESS 3 PRESS until the Sanitization Key programming screen appears 4 PRESS to delete all Sanitization keys 5 PRESENT the new Sanitization Key to the reader 6 PRESS to load new Sanitization...

Страница 16: ...filter replacement 2 PRESS to confirm that resetting is required OR PRESS to abort reset 3 PRESS Step 3 Composite vent filter replacement date 1 PRESS button to reset filter replacement date OR PRESS...

Страница 17: ...et 3 PRESS Step 6 Ultra Microfilter replacement date Optional 1 PRESS button to reset Ultra Microfilter replacement date OR PRESS to accept replacement date and proceed to Sanitization reminder 2 PRES...

Страница 18: ...this mode by PRESSING 4 2 Continuous Recirculation 24 7 screen 029 Refer to Section 3 2 Step 13 Your PURELAB 7000 US can be programmed to operate on specific days between selected times This is to op...

Страница 19: ...ew composite vent filter from packaging 5 WRITE the installation date on the filter label for future reference 6 INSTALL filter 7 SWITCH on power 8 RESET consumable reminder as described in Section 3...

Страница 20: ...doors and LOCATE Labpure L1 cartridge 3 PUSH Labpure L1 FORWARD 1 then LIFT 2 finally PULL 3 to remove and discard the cartridge 4 REMOVE new Labpure L1 from packaging 5 REMOVE sealing plugs from inle...

Страница 21: ...V LAMP WEAR CUT RESISTANT GLOVES MERCURY HAZARD DO NOT BREAK THE UV LAMP CONTAINS A SMALL AMOUNT OF MERCURY MAY CAUSE REDNESS OR IRRITATION AS A RESULT OF CONTACT WITH SKIN AND OR EYES 7 NOTE orientat...

Страница 22: ...ve lid from break tank and RELEASE any residual system pressure by PRESSING down the float valve 4 LIFT wing 1 and rotate anti clockwise 5 PIVOT latch 2 to release retaining arm 3 6 SWING retaining ar...

Страница 23: ...outlet tubing from the cartridge 4 REMOVE cartridge from clip and discard 5 REMOVE new E cartridge from packaging 6 REMOVE sealing plugs from inlet and outlet ports 7 REMOVE caps from cartridge side...

Страница 24: ...ATERIAL SAFETY DATA SHEET AND ANY APPLICABLE OSHA STANDARDS FOR THE CHEMICAL BEING USED WARNING LABEL THE MACHINE WITH APPROPRIATE WARNING SIGNS SUCH AS DO NOT USE CONTAINS STERILANT NOT PROVIDED 6 1...

Страница 25: ...sanitization 4 PRESS to enter RO sanitization Large white lid removal CAUTION ENSURE THE CORRECT DOSAGE OF CHEMICAL IS USED 5 UNSCREW lid on the left of the unit 6 Measure 50 ml of Minncare cold ster...

Страница 26: ...he step within the sanitization is complete the unit will go straight into process and continue to fill the recirculation reservoir Set sanitization reminder 10 PRESS to accept the sanitization remind...

Страница 27: ...screen 3 PRESS to enter sanitization process Enter RO sanitization 4 PRESS to show recirculation sanitization icon 5 PRESS sanitization to select recirculation Reservoir level adjustment 6 PRESS to al...

Страница 28: ...Minncare cold sterilant and slowly pour into the main reservoir 9 REPLACE lid Large white lid removal 10 UNSCREW lid on the left of the unit 11 Measure 50ml of Minncare cold sterilant and slowly pour...

Страница 29: ...complete the unit will go straight into process and continue to fill the recirculation reservoir Set sanitization reminder 15 PRESS to accept the sanitization reminder The system is now commissioned a...

Страница 30: ...ICABLE OSHA STANDARDS FOR THE CHEMICAL BEING USED WARNING LABEL THE MACHINE WITH APPROPRIATE WARNING SIGNS SUCH AS DO NOT USE CONTAINS STERILANT NOT PROVIDED 6 2 1 RO sanitization Refer to the EfferSa...

Страница 31: ...RO sanitization CAUTION ENSURE THE CORRECT DOSAGE OF CHEMICAL IS USED Large white lid removal 6 CUT 1 EfferSan tablet into quarters using tablet cutter 7 UNSCREW the large white lid on the left of the...

Страница 32: ...the step within the sanitization is complete the unit will go straight into process and continue to fill the recirculation reservoir Set sanitization reminder 12 PRESS to accept the sanitization remin...

Страница 33: ...screen 3 PRESS to enter sanitization process Enter RO sanitization 4 PRESS to show recirculation sanitization icon 5 PRESS sanitization to select recirculation Reservoir level adjustment 6 PRESS to al...

Страница 34: ...al CAUTION ENSURE THE CORRECT DOSAGE OF CHEMICAL IS USED 7 CUT 1 EfferSan tablet in half using tablet cutter 8 UNSCREW lid on the right of the unit 9 ADD of EfferSan tablets 10 REPLACE lid Large white...

Страница 35: ...B 7000 US MANU40676 Version 02 02 16 Page 33 Confirmation EfferSan tablet is added 15 PRESENT GREEN sanitization key to confirm that the EfferSan tablets have been added Note These two screens are sho...

Страница 36: ...complete the unit will go straight into process and continue to fill the recirculation reservoir Set sanitization reminder 17 PRESS to accept the sanitization reminder The system is now commissioned a...

Страница 37: ...dry place Wear protective clothing when handling Spillage If tablets are dry and uncontaminated collect up and place in heavy duty plastic bag Do not return to original container Wash away any residu...

Страница 38: ...urce of high temperature and rectify 672 94 Recirculation loop over temperature PRESS mute to silence alarm Water temperature is unacceptably high 50 C Contact Technical support 672 93 Protek incorrec...

Страница 39: ...Add chemical to break tank Option OFF Add chemical to reservoir Option ON Auto restart Output Bell Recirculate Calibration point Reset Cancel Sanitization key Clock Sanitization reminder Data Save dat...

Страница 40: ...2 Cloramine 1 ppm CL2 Silica 30 ppm SiO2 Fouling Index 10 FL Iron Manganese 0 1 ppm Fe Mn Organics 3 ppm TOC TEMPERATURE 4 40 C Recommended 15 25 C FLOWRATE maximum requirement at 15 C 9 l min Drain R...

Страница 41: ...er fail safe Water temperature alarm Water purity alarm Leak detection alarm Access restricted by User Recognition key Low voltage control circuit 24 V dc Visual alarms Audible alarms Output to operat...

Страница 42: ...CULAR PURPOSE The warranty provided herein and the data specifications and descriptions of the VWS UK Ltd systems appearing in VWS UK Ltd published catalogues and product literature may not be altered...

Страница 43: ...Minncare is a trademark of Minntech corporation EfferSan is a trademark of Activon Inc 2008 ELGA LabWater VWS UK Ltd All rights reserved Copyright Note The information contained in this document is t...