ELGA
MEDICA EDI
– US Operator Manual
MEDICA EDI - US MANU40685 Ver 2
– 10/15
Page 23
5.5
Replacing Protek L2 (LC177) or Carbon Filter
(LC123) Pre-treatment.
The replacement frequency of the pre-treatment cartridge is dictated
by the purity of the feed water. It should be replaced in the following
circumstances:
When indicated by the consumable alarm
When indicated by the alarm 75
After replacement of RO modules
5.5.1 Protek L2 (LC177)
1.
ENSURE process is OFF and ISOLATE power.
2.
OPEN front doors
3.
LOCATE Protek
4.
Remove lid from break tank and RELEASE any
residual system pressure by OPERATING float valve.
5.
TWIST to RELEASE clip and OPEN retaining clamp
at top of cartridge Protek L2 (LC177).
6.
PULL Protek L2 (LC177) forward
7.
If replacing a Protek, discard the used cartridge in
accordance with local regulations.
8.
REMOVE new cartridge from packaging.
9.
REMOVE sealing plugs from inlet and outlet ports.
10.
Wet ‘O’ rings and SLID Protek until it is fully engaged
11. CLOSE retaining clamp and fasten CLIP
12. The unit will automatically recognize the new cartridge
and date.
13. RESET reminder as described in section 3.3, Step 5.
Protek L2 (LC177)
Protek L2 (LC177)
Protek position
Protek L2