MEDICA EDI
–US MANU40685 Ver 2 – 10/15
5.4
Replacing UV lamp (LC105)
The UV lamp should be replaced in the following
circumstances:
When indicated by the consumable alarm.
After a maximum of twelve months use.
1.
ENSURE process is OFF and ISOLATE power.
2.
OPEN the front doors.
3.
REMOVE pack to improve access to the UV lamp
assembly
4.
LOCATE UV assembly on right-hand side of unit.
5.
REMOVE retaining spring clips from top and bottom of
the lamp assembly
6.
REMOVE electrical connectors from top and bottom
and take lamp from housing.
WARNING! IT IS STRONGLY RECOMMENDED THAT
DURING THE HANDLING OF THE LAMP
CUT-RESISTANT GLOVES BE WORN.
7.
DISCARD lamp.
Caution!
Dispose of lamp in accordance with local
authority regulations.
8.
REMOVE new lamp from packaging and follow
instructions included for cleaning.
9.
REFIT into UV assembly.
10.
RECONNECT to electrical connectors top and bottom.
11.
REFIT retaining spring clips top and bottom
12.
REPLACE Medpure pack (if fitted).
13.
RESET consumable reminder as described in Section
3.3, Step 4.
UV Lamp connection
UV lamp position
UV lamp