ELGA
MEDICA EDI
– US Operator Manual
MEDICA EDI - US MANU40685 Ver 2
– 10/15
Page 31
7.
EMERGENCY BYPASS
Contact your local service provider before
putting the unit into emergency bypass.
7.1
Emergency bypass
1.
Ensure the process is OFF
2.
Disconnect the electrical supply
3.
Open the doors.
4.
Close valve 1
5.
Close valve 2
6.
Remove yellow plastic cover 3
7.
Open valve 3
8.
Close valve 3 when there is no requirement for
purified water.
9.
Monitor water purity independently or install new
LC207 cartridges to guarantee water purity.
.
1
2
3