AVERAGE LIMITATION MODE
:
Output current is limited with linear trend during a specific time (t
AVG
) in order to reach the
limitation set. Typical limitation time (t
AVG
) is 200ms.
Limitations is enable both AC and DC mode
2.3.1.
LIMITATION SETTING
Both peak and average limitation can be set via serial command. Limits must be included in the
range [500,4095]. Values less than 500 are considered 500 by default (minimum value).
The factory values for both limitations are 4095. For further information please read the
programming manual.