4-2. Measurement Specifications

Accuracy of specifications apply for 1 year after purchase when
operated in a temperature of 18

O

C to 28

O

C (64

O

F to 82

O

F) and

a relative humidity of less than 75%.
Basic electrical specifications are given as +([% of reading] +
[number of least significant digits]).

DC Voltage

200mV

100

µ

V

+(0.25% of rdg+2dgt)

2000mV

1mV

+(0.5% of rdg+2dgt)

20V

10mV

+(0.5% of rdg+2dgt)

200V

100mV

+(0.5% of rdg+2dgt)

1000V

1V

+(0.8% of rdg+2dgt)

OVERLOAD PROTECTION: 250Vrms AC for 200mV range and 1000VDC
or 750Vrms AC for other ranges.

AC Voltage

200V

100mV

+(1.2% of rdg+10dgt)

750V

1V

+(1.2% of rdg+10dgt)

OVERLOAD PROTECTION: 1000VDC for 750Vrms for all ranges.

RESPONSE: Average responding, calibrated in rms of a sine wave.

FREQUENCY RANGE: 45Hz - 450Hz

DC Current

2000

µ

A

1

µ

A

+(1% of rdg+2dgt)

20mA

10

µ

A

+(1% of rdg+2dgt)

200mA

100

µ

A

+(1.5% of rdg+2dgt)

10A

10mA

+(3% of rdg+2dgt)

OVERLOAD PROTECTION: 250mA 250V fuse (10A range unfused).
MEASURING VOLTAGE DROP: 200mV.

AC Voltage Measurement

1. Plug the red test lead into the “V

mA” jack and

the black lead into the “COM” jack.

2. Set the RANGE switch to the desired ACV position.

3. Connect the test leads to the device or circuit being

measured.

4. Turn on the power of the device or circuit being

measured. The voltage value will appear on the
digital display.

DC Current Measurement

1. Plug the red test lead into the “V

mA” jack and

the black test lead into the “COM” jack (for
measurements between 200mA and 10A, connect
the red test lead to the “10A” jack. Make sure that
the test leads are pushed all the way into the jack).

2. Set the RANGE switch to the desired DCA position.

3. Open the circuit to be measured and connect the

test leads IN SERIES with the load in which current
is to be measured.

4. Read the current value on the digital display.

Range

Resolution

Accuracy

Range

Resolution

Accuracy

Range

Resolution

Accuracy

-6-

-11-

Содержание M-1000D

Страница 1: ...ectronics Inc 150 Carpenter Avenue Wheeling IL 60090 847 541 3800 Website www elenco com e mail elenco elenco com OPERATOR S INSTRUCTION MANUAL DIGITAL MULTIMETER M 1000D Copyright 2008 Elenco Electro...

Страница 2: ...screws from the back of the case 3 Remove the fuse from the holder and replace it with a new fuse of the specified value Fuse Numbers 250mA 250V Bussman Fuse GMA 250 250mA 250V Little Fuse 235 250 25...

Страница 3: ...off the power to the circuit under test before cutting unsoldering or breaking the circuit Small amounts of current can be dangerous Use caution when working above 60VDC or 30VAC rms Such voltages pos...

Страница 4: ...This marking indicates that equipment is protected completely be the double insulation 2 Introduction This meter is a 3 1 2 digit digital multimeter that is rugged reliable and convenient to use while...

Страница 5: ...the black test lead to the cathode 4 The forward voltage drop in mV will be displayed If the diode is reversed the number 1 will be displayed 3 Features 7 Functions 17 Measuring Ranges 0 25 Basic Acc...

Страница 6: ...dg 2dgt 10A 10mA 3 of rdg 2dgt OVERLOAD PROTECTION 250mA 250V fuse 10A range unfused MEASURING VOLTAGE DROP 200mV AC Voltage Measurement 1 Plug the red test lead into the V mA jack and the black lead...

Страница 7: ...3 Connect the test leads to the device or circuit being measured 4 Turn on the power of the device or circuit being measured The voltage value will appear on the digital display along with the voltage...

Страница 8: ...ing the unit in places with rapid temperature variations 4 In order to prevent damage or injury to the unit never fail to keep the maximum tolerable voltage and current especially for the 10A current...

Отзывы: