Gener a l info r ma tio n
E l e k t r o n E y e T e c h n o l o g y 2 0 1 8
4 .1 .3
Ref rac tiv e c orrec tion
It is important that the patient wears the correct refractive correction (suitable for a 25cm test
distance) during the visual field test.
Following the version 4.0 software update the Henson 9000 is now fitted with a full aperture trial
lens holder. Place the correct lenses in the holder and with the patient correctly aligned move
the holder as close to the patients eye as possible. Earlier Henson 9000s were designed to be
used with a special perimetric lens set and head mounted frame. The integral lens holder is
intended to be an improvement on this solution. [Note: This holder is available to retro-fit
should you require one for your instrument].
An occluder is placed in front of the eye not being tested.
Recommended additions (lens power to be added to the patient's current distance prescription)
are given in the table below.
Patient's age
Add
40-44
+1.50
45-49
+2.00
50-54
+2.50
55-59
+3.00
60-64
+3.50
>65
+4.00