Appendix 3 - Installation
1. LOCATION
The Henson should be placed on a suitable flat surface, with no part of the unit over hanging
the edge of the surface.
The patient aperture should be positioned away from direct light.
The electrical installation of the room where the Henson Perimeter is to be operated must
comply with the "Regulation for the electrical equipment of buildings" published by the
Institution of Electrical Engineers. The unit and any ancillary parts must be protected from
ingress of liquids and flammable anaesthetic mixtures.
2. MAINS SUPPLY
The mains supply required is 250 VA (maximum, depending on instrument) 110-240Vac.
The supplied mains lead must be used and, if faulty, must be replaced by an IEC approved*
mains lead with conductors of at least 0.75mm
2
cross sectional area. Replacement mains leads
are available from the manufacturer, see the
section.
Access to the mains supply plug and socket must be maintained at all times as this is the means
of isolating the device from the mains supply.
*For USA/Canada - The supplied Hospital grade mains cable must be used but grounding
reliability can only be achieved when the equipment is connected to an equivalent receptacle
marked "Hospital Only" or "Hospital Grade". Outside of the hospital environment the Hospital
grade mains cable must be used and connected to a grounded (Earthed) outlet to maintain
grounding of the equipment.
3. INTE RCONNE CTING
Connect the supplied mains lead from the mains input connector on the instrument to the
mains supply (either direct from the wall socket or via the electric table).
Plug the patient response button jack plug into the socket on the rear panel of the unit.
4. ACCE SSORIE S
If the unit is located on a table with in-built electrical connections, the table should be
connected to the mains supply using a suitable mains connection lead and the Henson should
be powered from the power outlet at the top of the table.
5. SAFE TY
Ensure the connected leads do not trail on the floor and are not subject to abrasion on sharp
edges. Use only computers, printers, computers and monitors that conform to EN60950.
6. E MC
The Henson range of perimeters conform to European Directive 2004/108 for EMC. If emitted
radiation causes interference with other items of equipment, position it further away or try a
different orientation. Do not operate transmitters or mobile telephones in close proximity to the
equipment.
5.2.2 Technical description
135
Содержание Henson 9000
Страница 1: ...Elektron Eye Technology Ltd Henson 9000 ...
Страница 49: ...49 Visual field tests E l e k t r o n E ye Te c h n o l o g y 2 0 1 8 ...
Страница 85: ...85 Options program E l e k t r o n E ye Te c h n o l o g y 2 0 1 8 ...
Страница 123: ...123 Appendix 3 Installation E l e k t r o n E ye Te c h n o l o g y 2 0 1 8 ...
Страница 124: ...124 Henson 9000 Perimeter User Manual E l e k t r o n E ye Te c h n o l o g y 2 0 1 8 ...
Страница 131: ...131 Appendix 4 Maintenance and warranty E l e k t r o n E ye Te c h n o l o g y 2 0 1 8 ...
Страница 152: ......