The settings of the various functions are described below. After setting it up, the function can be loaded.
►
How to load a function
1.
After setting the values for the required signal generation, tap on the
touch area
.
The device will then load the data into the internal controller and changes the
display. Shortly afterwards the static values are set (power and voltage or cur-
rent), the DC output is switched on and the touch area
enabled.
Only then can the function be started.
The static values are applied to the DC output immediately after loading the function, because it
switches the DC output on automatically in order to settle the start situation. These static values
represent start and end values for the progress of the function, so that the function does not
need to start from 0. Only exception: when applying any function to the current (I), there is no
adjustable static current value, so the function would always start from 0 A.
►
How to start and stop a function
1.
The function can be
started
either by tapping
or pushing the “On/Off” button, if the DC output
is currently switched off. The function would then start immediately.
2.
The function can be
stop
ped either by tapping
or operating the “On/Off” button. However, there
is a difference:
a) The
button stops only the function, the DC output remains ON with the static values.
b) The “On/Off“ button stops the function and switches off the DC output.
Every device alarm (overvoltage, overtemperature etc.), protection (OPP, OCP) or event with
action = Alarm stops the function progress automatically, switches off the DC output and reports
the alarm.
3.10.5
Sine wave function
The following parameters can be configured for a sine wave function:
Value
Range
Description
I(A), U(A)
0...(Nominal value - (Offs)) of U, I
A = Amplitude of the signal to be generated
I(Offs), U(Offs)
(A)...(Nominal value - (A)) of U, I
Offs = Offset, based on the zero point of the mathematical
sine curve, may not be smaller than the amplitude.
f (1/t)
1...10000 Hz
Static frequency of the signal to be generated
Schematic diagram:
Application and result:
O
ffs
A
m
plit
ud
e
t
U,I
f
A
m
plit
ud
e
A normal sine wave signal is generated and applied to
the selected set value, e. g. voltage (U). At a constant
load resistance, the output voltage and thus also the
output current will follow a sine wave.
For calculating the maximum power output the amplitude
and offset values for the current have to be added.
Example: an output voltage of 30 V is set together with
sin(I) with an amplitude of 12 A and an offset of 15 A.
The resulting maximum power output is then achieved
at the highest point of the sine wave and is (12 A + 15
A) * 30 V = 810 W.