PSI 9000 DT Series
1.8.3
Specific technical data
(1 Related to the nominal values, the accuracy defines the maximum deviation between an adjusted values and the true (actual) value.
Example: a 80 V model has min. 0.1% voltage accuracy, that is 80 mV. When adjusting the voltage to 5 V, the actual value is allowed to differ max. 80 mV, which
means it might be between 4.92 V and 5.08 V.
(2 RMS value: LF 0...300 kHz, PP value: HF 0...20 MHz
(3 The display error adds to the error of the related actual value on the DC output
320 W
Model
PSI 9040-20 DT
PSI 9080-10 DT
PSI 9200-04 DT
AC Input
Voltage range
90...264 V AC
90...264 V AC
90...264 V AC
Connection
1ph,N,PE
1ph,N,PE
1ph,N,PE
Frequency
45-65 Hz
45-65 Hz
45-65 Hz
Fusing
MT 8 A
MT 8 A
MT 8 A
Leak current
< 3.5 mA
< 3.5 mA
< 3.5 mA
Power factor
≈ 0.99
≈ 0.99
≈ 0.99
DC Output
Max. output voltage U
Max
40.8 V
81.6 V
204 V
Max. output current I
Max
20 A
10 A
4 A
Max. output power P
Max
320 W
320 W
320 W
Overvoltage protection range
0...44 V
0...88 V
0...220 V
Overcurrent protection range
0...22 A
0...11 A
0...4.4 A
Overpower protection range
0…352 W
0…352 W
0…352 W
Output capacitance
4760 μF
3400 μF
720 μF
Temperature coefficient for set
values Δ/K
Voltage / current: 100 ppm
Voltage regulation
Adjustment range
0...40.8 V
0...81.6 V
0...204 V
Accuracy
(1
(at 23 ± 5°C)
< 0.1% U
Max
< 0.1% U
Max
< 0.1% U
Max
Line regulation at ±10% ΔU
AC
< 0.02% U
Max
< 0.02% U
Max
< 0.02% U
Max
Load regulation at 0...100% load
< 0.05% U
Max
< 0.05% U
Max
< 0.05% U
Max
Rise time 0...100% (@full load)
Max. 30 ms
Max. 60 ms
Max. 65 ms
Settling time after load step
< 1.5 ms
< 1.5 ms
< 1.5 ms
Display: Resolution
„1.9.5.4. Resolution of the displayed values“
Display: Accuracy
(3
≤ 0.2% U
Max
≤ 0.2% U
Max
≤ 0.2% U
Max
Ripple
(2
< 8 mV
PP
< 0.8 mV
RMS
< 8 mV
PP
< 0.8 mV
RMS
< 20 mV
PP
< 2.5 mV
RMS
Remote sensing compensation
Max. 5% U
Max
Max. 5% U
Max
Max. 5% U
Max
Output voltage fall time (at no load)
after switching DC output off
-
Down from 100% to <60 V: less than 10 s
Current regulation
Adjustment range
0...20,4 A
0...10,2 A
0...4,08 A
Accuracy
(1
(at 23 ± 5°C)
< 0.2% I
Max
< 0.2% I
Max
< 0.2% I
Max
Line regulation at ±10% ΔU
AC
< 0.05% I
Max
< 0.05% I
Max
< 0.05% I
Max
Load regulation at 0...100% ΔU
OUT
< 0.15% I
Max
< 0.15% I
Max
< 0.15% I
Max
Ripple
(2
< 1 mA
RMS
< 1 mA
RMS
< 1,5 mA
RMS
Display: Resolution
„1.9.5.4. Resolution of the displayed values“
Display: Accuracy
(3
≤ 0.2% I
Max
≤ 0.2% I
Max
≤ 0.2% I
Max
Power regulation
Adjustment range
0…326.4 W
0…326.4 W
0…326.4 W
Accuracy
(1
(at 23 ± 5°C)
< 1% P
Max
< 1% P
Max
< 1% P
Max
Line regulation at ±10% ΔU
AC
< 0.05% P
Max
< 0.05% P
Max
< 0.05% P
Max
Load reg. at 10-90% ΔU
OUT
* ΔI
OUT
< 0.75% P
Max
< 0.75% P
Max
< 0.75% P
Max