EL 9000 B 15U/24U Series
IF Level 1
Description
Profibus DP
Node Address
Selection of the Profibus or node address of the device within range 1...125 via direct input
Function Tag
String input box for a user-definable text which describes the Profibus slave function tag.
Max. length: 32 characters
Location Tag
String input box for a user-definable text which describes the Profibus slave location tag.
Max. length: 22 characters
Installation Date
String input box for a user-definable text which describes the Profibus slave installation
date tag. Max. length: 40 characters
Description
String input box for a user-definable text which describes the Profibus slave.
Max. length: 54 characters
IF Level 1
Level 2
Description
CANopen
Node Address
Selection of the CANopen node address in the range 1...127
Baud Rate
AUTO
Automatic detection of the bus baud rate or speed
LSS
Automatically sets baud rate and node address
Manual
Manual selection of the baud rate that is used by the CANopen interface.
Possible selections: 10 kbps, 20 kbps, 50 kbps, 100 kbps, 125 kbps, 250
kbps, 500 kbps, 800 kbps, 1Mbps (1Mbps = 1 Mbit/s, 10 kbps = 10 kbit/s)
IF Level 1
Level 2 Level 3
Description
Ethernet / ModBus-TCP
, 1 & 2 Port
IP Settings 1
DHCP
The IF allows a DHCP server to allocate an IP address, a subnet
mask and a gateway. If no DHCP server is in the network then
network parameters will be set as defined in item “Manual”
Manual IP address
This option is activated by default. An IP address can be man-
ually allocated.
Gateway
Here a gateway address can be allocated if required..
Subnet mask
Here a subnet mask can be defined if the default subnet
mask isn’t suitable.
DNS address 1
DNS address 2
Here the addresses of the first and second Domain Name Serv-
ers (DNS) can be defined, if needed. A DNS is only necessary
if the device shall be registered with its domain and host name,
in order to access it in a browser more comfortably.
Port
Range: 0...65535. Default ports:
5025
= Modbus RTU (all Ethernet interfaces)
Reserved ports which must not be set with this parameter:
502
= Modbus TCP (Modbus-TCP interface only)
other typical reserved ports
IP Settings 2-P1
IP Settings 2-P2
AUTO
Settings for the Ethernet port such as transmission speed are
set automatically.
Manual Half duplex
Manual selection for transmission speed (10MBit/100MBit) and
duplex mode (full/half). It’s recommended to use the “AUTO”
option and only revert to “Manual” if these parameters fail.
Full duplex
10MBit
100MBit
Host name
Free choice of host name (default: Client)
Domain name
Free choice of Domain (default: Workgroup)
TCP Keep-Alive Enable TCP keep-alive
IF Level 1
Description
RS232
-
The baud rate is selectable, other serial settings can’t be changed and are defined like
this: 8 data bits, 1 stop bit, parity = none
Baud rates: 2400, 4800, 9600, 19200, 38400, 57600, 115200