Do not position the device in
such a way that it would be
difficult to reach the main power
source and that might prevent
the rapid closure of the device
if needed.
THE PATEINT IS THE
DESIGNATED OPERATOR
The patient can use the buttons and change the battery under normal conditions and
maintain the device and its accessories according to the user manual
ATTENTION!
•
Use only Electro-Medic rechargeable Li-ion batteries
•
NEVER reverse the (+) and (-) terminals when connecting them or let the batteries
come into contact with metal objects (necklaces, hairpins, etc.)
•
NEVER charge Li-ion batteries for longer than 72 hours
•
The battery charger must be in compliance with IEC 60601-1 standards
SAFETY MEASURES
•
Do not expose the equipment to fire, direct sunlight or other heat sources, which may
cause a fire or explosion, or generate toxic gases
•
Do not store or transport the device with metal objects
•
Do not disassemble or modify the device components
•
Avoid all contact with water or any other liquid
INSTRUCTIONS FOR USING THE CHARGER
•
Insert a Li-ion battery.
Align the connection terminals (+) and (-) correctly
•
Plug the charger into a standard wall outlet
•
A red or yellow LED indicates charging
•
When fully charged, the LED turns green. Unplug the charger and remove the battery
CHARGING TIME
•
A Li-ion battery takes about 3.5 hours to charge.
LI-ION BATTERY
•
Limited voltage: 4.2 V
•
Rechargeable Li-ion battery: 3.7 V/600 mAh
ADAPTOR
•
100-240 v 50-60 Hz, 1.2A
ATTENTION
•
Connecting this equipment to an adaptor other than the one provided with the
Electro-Medic equipment is not allowed.
SAFETY MEASURES
•
Do not cause a short-circuit
•
Do not expose the device to high temperatures
•
Use the charger only as specifically recommended
The operator’s
immediate response is
required
Colours of indicator lights
and their meaning
The operator’s rapid
response is required
Ready to use
Meaning other than
those here
green
yellow
Other
red
The adaptor is a 2MOPP piece of equipment under
IEC 60601-1-1.
Approval of the equipment is valid if used in
combination with the adaptor provided with this
equipment.
Type BF applied part:
Electrodes
IP22
Continuous operation
Internally powered
Electro-Medic device
Applied part
Electrical equipment
protected against harmful
ingress of water or
particulate matter.
Operation mode
Note: Not intended to be sterilized.
Not for use in an oxygen-rich environment.
Protection from
electric shock
SAFETY CLASSIFICATION OF
ELECTRO-MEDIC EQUIPMENT