22
PROGRAMMING
MUSCLE STIMULATOR
NMES
( SM )
1
Press the ON/OFF button.
Press the PROGRAM button OR to go to the following program or return to the previous program until the program
is at P28-P30.
Press the
S
(SELECT) button and hold it down for 3 seconds in order to enter programming mode.
(Please refer to the programming chart for the next steps.)
FIRST STEP
CHANNEL #1
The first step is to choose between:
SM
: NMES Muscle Stimulator
or
T
: TENS
Press the INCREASE (or DECREASE) button to go from
SM
to
T
and vice versa.
Confirm your selection by pressing the
S
(SELECT) button
SECOND STEP
Press the INCREASE (or DECREASE) button to display the different types of stimulations available on the screen.
The programming chart (6.0) indicates the different settings that are possible for each type of stimulation.
Confirm your selection at each step by pressing the
S
(SELECT) button
Types of stimulation for the TENS mode
• Conventional (
C
)
• Burst (
B
)
• Pulse duration modulation (
MW
)
• Pulse frequency modulation (
MR
)
THIRD STEP
CHANNEL #2
• Repeat the previous steps for channel #2
Confirm your selection at each step by pressing the
S
(SELECT) button
PROGRAMMING
TENS
(T)
INDEPENDENT CHANNELS
2
3
PR
+
PR
-
5.2
23
page