Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
gammaCore is intended for use in the electromagnetic environment specified below.
The customer or the user of gammaCore should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
Environment - Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 2 kV for AC Mains
± 1 kV Other
Mains power quality should be
that of a typical commercial
and/or hospital environment.
Surge
IEC 61000-4-5
± 1 kV for power
supply lines
± 2 kV line(s)
to earth
± 1 kV for power
supply lines
± 2 kV line(s)
to earth
Mains power quality should be
that of a typical commercial
and/or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5%
U
T
(>95% dip in
U
T
)
for 0.5 cycles
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
<5%
U
T
(95% dip in
U
T
)
for 5 cycles
<5%
U
T
(>95% dip in
U
T
)
for 0.5 cycles
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
<5%
U
T
(95% dip in
U
T
)
for 5 cycles
Mains power quality
should be that of a typical
commercial and/or hospital
environment. If the user of the
product requires continued
operation during power mains
interruptions, it is recommended
that the product be powered
from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
Содержание gammaCore Sapphire SLC
Страница 26: ...26 64000 00154 Rev1 26 64000 00154 Rev 1 ADDITIONAL INFORMATION FOR HEALTHCARE PROVIDERS ...
Страница 35: ...35 64000 00154 Rev1 Notes ...
Страница 36: ...36 64000 00154 Rev1 Notes ...
Страница 37: ...electroCore Inc 200 Forge Way Suite 205 Rockaway NJ 07866 United States www electrocore com ...
Страница 74: ...74 64000 00154 Rev1 Notes ...
Страница 75: ...75 64000 00154 Rev1 Notes ...
Страница 76: ...electroCore Inc 200 Forge Way Suite 205 Rockaway NJ 07866 États Unis www electrocore com ...