14
64000-00154 Rev1
10. RELOADING INSTRUCTIONS
1. Turn gammaCore on by pressing the power button.
Have your refill card ready.
2. Once the device is turned on, immediately place the
refill card across the device so you can see the display
(as shown in the diagram).
3. gammaCore will display “rd” and the refill icon when
gammaCore is reading the refill card.
4. gammaCore will
beep twice when the device has
been loaded with months/days (Refer to Section 8 for
display details). This may take several seconds.
The device is now ready to be used for treatment.
NOTE: If “bd” is displayed, an error has occurred during
the loading process. Turn device off and try again.
NOTE: If 5 consecutive loading errors occur, “Ec” will
be displayed. Wait 24 hours and retry, or contact
Customer Service.