Elcometer PG70ABDL Скачать руководство пользователя страница 16

15

en

8.1 BEFORE YOU START

Prepare the surface.

Ensure the correct transducer is selected and set in the gauge.

Select the correct measurement mode - see “Measurement - Modes” on page 10.

Set the zero point of the transducer,if necesssary - see:

“Transducer - Zeroing” on page 12 and

Calibrate the gauge - see:

“Calibrating - for a Material” on page 13 and

Choose the measurement view (DIGITS, RF or RECT, or B-SCAN) - see “The Measurement 
Screen” on page 7.

8.2 PROCEDURE
1.

Apply couplant

For the gauge to work correctly there must be no air gap between the transducer and the 
surface of the material to be measured. This is achieved using a couplant.
Before the transducer is placed on the surface, put a small amount of couplant supplied with 
the gauge on the surface of the material. Typically a single drop is sufficient.

2.

Place transducer onto the surface of the material to be measured

Press the transducer wearface into the couplant. Only moderate pressure  is necessary to 
keep the transducer stationary and the wearface seated flat against the surface of the 
material.

3.

Read display

If six or seven bars of the stability indicator are showing, the display will be reading the correct 
thickness of the material directly beneath the transducer.
If the stability indicator has fewer than five bars showing, or the numbers on the display seem 
erratic, check to make sure that there is an adequate film of couplant beneath the transducer, 
and that the transducer is seated flat against the material. If the condition persists, it may be 
necessary to:

adjust the gain, or

select a different transducer (size or frequency) for the material being measured.

The gauge will perform a number of measurements every second when the transducer is in 
contact with the surface of the material. The display is updated as each reading is taken.

4.

Remove transducer from surface

The display will show the last measurement made.

Note: Occasionally, a small film of couplant will be drawn out between the transducer and the
surface as the transducer is removed. When this happens, the gauge may perform a measurement
through this couplant film. This phenomenon can be seen when one thickness value is observed
while the transducer is in place, and another value is observed after the transducer is removed. If
this happens, take the reading again using less couplant.

9 GATES

9.1 INTRODUCTION

The Model PG70ABDL is equipped with gates which control the time measurement process
allowing you to measure a specific region on or between waveforms.
Further information on gates may be found on the Elcometer NDT Knowledge Centre on
www.elcometerndt.com.

Содержание PG70ABDL

Страница 1: ...0 en Model PG70ABDL Precision Gauge Operating Instructions...

Страница 2: ...re available to download via our website Elcometer NDT Ultrasonic Couplant Material Safety Data Sheet www elcometerndt com images MSDS elcometer_ultrasonic_couplant pdf Elcometer NDT Ultrasonic Coupla...

Страница 3: ...ing Readings 14 9 Gates 15 10 Thru Paint Measurement Technique 16 11 Measurement Options 16 12 Gauge Setups 20 13 Data Transfer Software 22 14 Storage 22 15 Maintenance 23 16 Technical Specification 2...

Страница 4: ...measured The Elcometer PG70ABDL includes a data logging memory facility which allows readings to be stored in batches before being downloaded to a computer 1 1 STANDARDS Your gauge can be used in acco...

Страница 5: ...a change or selection If your gauge is displaying a grid log this key toggles an advance to row number option In the MENU MEAS and EDIT functions this key acts as a back or escape function If your gau...

Страница 6: ...apply couplant to the connecting end of the delay line before screwing together The transducer transmits and receives ultrasonic sound waves that the gauge uses to calculate the thickness of the mater...

Страница 7: ...displayed in English Spanish and German 1 Select MENU SETUP LANGUAGE and adjust with LEFT and RIGHT arrows Press once to access the menus then press to scroll to the right along the sub menu bar and t...

Страница 8: ...asurement screens A Battery Icon Remaining battery life B Repeatability Stability Indicator This indicator is used in conjunction with the digital thickness values displayed When all the vertical bars...

Страница 9: ...surface It can also be a useful view when scanning for pits and flaws The B Scan display is equipped with a scan bar representing the overall thickness The scan bar gives the user a visual indication...

Страница 10: ...value 3 Press MEAS to return to the measurement 5 7 GAIN The gain the amplitude of the transmitted pulse can be adjusted to suit a variety of applications To obtain valid readings the gain must be se...

Страница 11: ...dge Centre on www elcometerndt com 5 10 FRONT PANEL LIGHTS The green light illuminates when the alarm mode is active and the measured thickness is within limits The red light illuminates when the alar...

Страница 12: ...nt range and the back wall echo is less likely to be lost in the ring down or noise from the interface echo The pictures show typical A scans when measuring in interface to echo and echo to echo respe...

Страница 13: ...ucer type with the newly selected transducer type You now need to perform a probe zero before taking measurements 7 2 TRANSDUCER ZEROING Note This function is only used when using a contact style tran...

Страница 14: ...n uses two samples of different thicknesses with enhanced accuracy over small ranges close to the sample thicknesses used For the highest measurement accuracy use the known thickness calibration as th...

Страница 15: ...and DOWN arrows 4 Press OK to calculate the velocity and return to the measurement screen or ESC to cancel Note Place the transducer back on the calibration point The thickness reading should now matc...

Страница 16: ...ly beneath the transducer If the stability indicator has fewer than five bars showing or the numbers on the display seem erratic check to make sure that there is an adequate film of couplant beneath t...

Страница 17: ...elect MENU UTIL AUTO FIND 11 2 HIGH SPEED SCAN Although your gauge excels at making single point measurements it is sometimes desirable to examine a larger region searching for the thinnest point The...

Страница 18: ...ption of displaying the waveform filled in black or outlined unfilled To set this option 1 Press MENU select DISP select VIEW and then select RECT to set the RECT display view 2 Press MENU scroll to t...

Страница 19: ...displayed along with the grid name on screen 11 8 HANDLING LOGGER FILES Once created logger files may be opened or closed using the functions OPEN and CLOSE Closing logger files places the gauge in i...

Страница 20: ...scroll to the cell location press CLR and then OK or ESC to cancel The reading is deleted and if you wish you can now take another measurement and save the reading in this cell location 11 11 EDITING...

Страница 21: ...e software includes a default setup file that can be uploaded to the gauge at any time to restore factory settings However you should consider saving modified setups to an empty location rather than o...

Страница 22: ...ss ENTER The setup list screen is displayed 5 Scroll through the list of setups until the required location to save the setup is highlighted 6 Press OK to activate the confirmation screen 7 Press OK t...

Страница 23: ...with the latest version of firmware 13 1 2 USB to Serial Converter Some newer laptop computers do not have standard serial ports In this case use the USB to Serial converter supplied with the gauge 1...

Страница 24: ...ometer NDT supplier or directly to Elcometer NDT The warranty will be invalidated if the instrument has been opened 15 2 TRANSDUCER The transducer will wear with repeated use Transducer life depends o...

Страница 25: ...in s Thickness 0 001 mm 0 0001 or 0 01 mm 0 001 selectable Velocity Calibration Range 1250 m s to 10000 m s 0 0492 in s to 0 3937 in s Weight including batteries 380 g 13 5 oz Dimensions W x H x D 63...

Страница 26: ...s you specific legal rights and you may also have other rights which may vary from country to country state to state or province to province 17 2 OBTAINING SERVICE DURING WARRANTY PERIOD If your hardw...

Страница 27: ...ep 2 5 25mm TW 24007 Replace with S1018 1018 Steel Block A Aluminium Block SS Stainless Steel Block T Titanium Block 18 3 ULTRASONIC COUPLANT Each gauge is supplied with a 120ml 4oz bottle of standard...

Страница 28: ...5232 0 206 PVC 2388 0 094 Copper 4674 0 184 Quartz Glass 5639 0 222 Epoxy Resin 2540 0 100 Approx Rubber Vulcanised 2311 0 091 German Silver 4750 0 187 Silver 3607 0 142 Glass Crown 5664 0 223 Steel...

Страница 29: ...ONE POINT Performs a single point calibration This function allows you to automatically calculate the velocity by entering a known sample thickness TWO POINT Performs a two point calibration This func...

Страница 30: ...cho GATE 1 Set the value of Gate 1 Gate 1 can be used in pulse echo echo echo and interface echo measurement modes GATE 2 Set the value of Gate 2 Gate 2 can be used in echo echo and interface echo mea...

Страница 31: ...is value a red light will illuminate and the alarm will sound XFER TRANSFER BACKUP SET UPS Allows you to backup the setups currently stored in your gauge to a computer via the RS232 port RESTORE SETUP...

Отзывы: