![Elcometer PG70ABDL Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/elcometer/pg70abdl/pg70abdl_operating-instructions-manual_2377275030.webp)
29
TUNE
GAIN
Increases or decreases the overall amplitude of the signal.
Gain is similar to the volume control on a stereo receiver.
AGC
When operating in echo-echo mode, the gauge adjusts the
gain automatically. Alternatively, the AGC can be manually
controlled.
THRESHOLD
Adjust the sensitivity level of the gauge. The amplitude of
the signal must reach and exceed the threshold level before
a measurement is detected.
POLARITY
The gauge operates on a zero crossing detection principle.
This feature toggles which stroke of the cycle the crossing
detection uses, either positive or negative.
RECT WAVE-
FORM
Toggles between an outlined or filled view option when the
display setting is in RECT (rectified) wave mode.
GATE
MEASURE
MODE
Select the measurement mode for different application
requirements. The modes are P-E (pulse-echo), E-E
(echo-echo) and I-E (interface-echo).
GATE 1
Set the value of Gate 1. Gate 1 can be used in pulse-echo,
echo-echo and interface-echo measurement modes.
GATE 2
Set the value of Gate 2. Gate 2 can be used in echo-echo
and interface-echo measurement modes only.
SETUP
OPEN
Displays a list of factory and user-defined setups currently
stored in memory. These setups can be recalled and used
at any time.
SAVE
Save a custom setup that you have modified or created.
DEFAULT
SETUP
Loads a basic default setup. Use only as a last resort if the
setups in your gauge have been corrupted and a computer
is not accessible.
LANGUAGE
Select the menu language
DATA
NEW
Creates a new alpha numeric grid file with auto identifiers.
You can customise the file according to your requirements.
EDIT
Modify the parameters of grid files previously saved.
Note: Predefined coordinates cannot be changed once they
have been created.
OPEN
Recalls existing grid files from gauge memory.
DELETE ONE
GRID
Deletes a single file from the gauge memory.
DELETE ALL
GRIDS
Deletes all files from the gauge memory.
Menu
Function
Description