Elbur PB 636 IV Скачать руководство пользователя страница 1

INSTRUCTION FOR USE

 

ELBUR Nursing Beds:           

PB 636 IV, 
PB 636 IV with axional brake

 

 

 

 

 

 

 
 

 

 

Содержание PB 636 IV

Страница 1: ...INSTRUCTION FOR USE ELBUR Nursing Beds PB 636 IV PB 636 IV with axional brake...

Страница 2: ......

Страница 3: ...er to meet customer expectations we provide wide range of bed height adjustment We also care about attractive design as well as matching furniture systems and accessories which can be chosen depending...

Страница 4: ...s and warning messages 10 2 4 1 General information 10 2 4 2 Electrical components information 11 2 3 3 Information for care personnel 12 2 3 4 Assembly information 13 2 3 5 Service information 13 3 0...

Страница 5: ...Assembly of side rails 21 6 12 Assembly of the lifting pole 22 7 0 Operating of the nursing bed 22 7 1 Electric mattress support adjustment 22 7 2 Side rails operating 24 7 3 Setting the springiness...

Страница 6: ...Medical electrical equipment Part 1 General requirements for basic safety and essential performance EN 60601 2 52 Medical electrical equipment Part 2 52 Particular requirements for basic safety and e...

Страница 7: ...e density of foamed material 35 50 kg m3 Drive system Drive lifting unit Linak LA40 Mattress support platform drive unit Linak LA27 Control box Linak CA40 Hand switch HL74 Input voltage 230V AC 50Hz O...

Страница 8: ...ontinuous use followed by 18 minutes not in use manufacturer 1 3 2 Information labelling Max Patient Weight 135 kg Safe Working Load 175 kg Dimensions of the standard mattress H x W x L 10 x 90 x 200...

Страница 9: ...ead instruction for use carefully and follow exactly included rules Declaration of Conformity manufacturer maximum patient weight safe working load 2 0 Safety information 2 1 Definitions of involved p...

Страница 10: ...ceeded It is recommended that patients taller than 190 cm should not use ELBUR nursing bed without a special extended bed frame which can be ordered from manufacturer It provides comfort of use withou...

Страница 11: ...patient falls Make sure that other attached equipment can function in a secure and safe manner do not hesitate to contact manufacturer with questions Make sure that all castors have their brakes appli...

Страница 12: ...onents must not be cleaned with a water jet or a high pressure cleaner Clean only with a moist towel 2 3 3 Information for care personnel Warnings Do not perform medical operation on the nursing bed D...

Страница 13: ...stment It cannot be blocked by any obstacles bedside lockers or windowsills Make sure that cables connected control box with hand switch and other drive units are not damaged that there is no loop in...

Страница 14: ...the nursing bed please read instruction manual carefully and verify the shipment contents with your order items described below may not be included in your model Please contact your distributor or El...

Страница 15: ...g sections Both of them are fitted under the bed in that case assembly is necessary see subclauses 6 4 6 6 4 3 Control box Control box CA40 is fitted under the mattress support platform in head sectio...

Страница 16: ...nursing bed to make sure that it is going to work properly Modification of nursing bed without authorization of Elbur company can caused hazards for patients ad operators for example entrapment betwee...

Страница 17: ...n danger situation there must be enough space available for adjusting the bed the space under the bed must remain clear there must be enough space for operator or caregiver to look after patient at le...

Страница 18: ...ort platform slowly and put it horizontal on the scissor lift align holes in the mattress support platform scissor lift lugs then connect it using quick release bolts 6 4 Location of drive units and c...

Страница 19: ...eath the mattress support platform align holes in holders and the drive insert the quick release bolt and secure by closing the retainer align holes in the second holder and the drive on the other its...

Страница 20: ...t insert blind stoper battery charging symbol unused connection slot insert blind stopper 3 mattress support platform drive unit adjustment of foot rest section 4 lifting drive unit up and down mattre...

Страница 21: ...s attached with end boards into tubes of the mattress support platform to the limit secure end boards and mattress support platform connection by screwing in and tightening the knurled screws 6 11 Ass...

Страница 22: ...of the bed 7 0 Operating of the nursing bed Before first use of the nursing bed please read instruction carefully especially potential risks described in subclause 2 3 which can appeared if safety pr...

Страница 23: ...e function the green light will appear which means function has been engaged again When the thigh rest is fixed the lower leg rest can be adjusted individually there are two adjusting ratchets which l...

Страница 24: ...e locking button and lower them slowly do not let to rapid decrease 7 3 Setting the springiness of mattress support platform Warning Rashness can caused finger entrapment between mattress support plat...

Страница 25: ...tion of castors unlock brakes on castors or axional brake nursing bed is ready to move After moving lock all castors plug in the mains power cable power supply cord shall be kept clear of any moving n...

Страница 26: ...h a moist towel Never use scouring agents abrasive cleaning agents stainless steel cleaners or scouring pads because they can damage the surface Before beginning the cleaning unplug the power cable fr...

Страница 27: ...u the subsequent technical safety check has to be documented It is obligated with every maintenance to perform regular inspections to ensure the safe condition of the nursing bed If damage or malfunct...

Страница 28: ...cturer s final inspection Electrical appliance class I II Manufacturer Elbur sp z o o sp k Bed type PB _________________________________________________ Serial number _________________________________...

Страница 29: ...operly YES NO Is lifting pole distorted YES NO Is triangle grab handle still good enough to use YES NO 1 2 Visual inspection of the electrical components Is the power supply cord connected properly an...

Страница 30: ...switch working properly YES NO Are drive units working properly abnormal noises smooth operation YES NO 3 2 Performance check of the mechanical components Is suspension system working properly smooth...

Страница 31: ..._________ ____________________________________________________________ Comments ____________________________________________________________ ___________________________________________________________...

Страница 32: ...ted Deactivate disabler turn key to the left Hand control damaged Notify customer service Control box damaged Notify customer service Lower leg rest section optional function does not adjust Adjusting...

Страница 33: ...nce of operating and storage conditions for ELBUR nursing beds Nursing bed shall be protected from destruction under external influences before re storage Disassembly of nursing bed shall be performed...

Страница 34: ...__________________________________________________________ ____________________________________________________________ ____________________________________________________________ ___________________...

Страница 35: ...PB 636 IV Elbur sp z o o sp k 35 of 36...

Страница 36: ...Elbur sp z o o sp k Dzia osza 34 56 500 Syc w POLAND Tel 48 62 786 97 80 Fax 48 62 786 97 81 e mail info elbur eu www elbur eu E TF 3E 13 01 23 11 2020...

Отзывы: